Z-0517-2020Class II
ASI 2.0 Calcar Trimmer Shaft for Use with Biomet Rasp, Model Number 110032331
Zimmer GmbH / initiated 2019-08-06 / Terminated
Zimmer Biomet, Inc. began a recall of ASI 2.0 Calcar Trimmer Shaft for Use with Biomet Rasp, Model Number 110032331 on . FDA classified it as Class II and gives the reason as "The potential that the end of the shaft could fail to effectively mate with the broach with the Calcar Planar Shaft instrument, part number 110032331". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0517-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- ASI 2.0 Calcar Trimmer Shaft for Use with Biomet Rasp, Model Number 110032331
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- The potential that the end of the shaft could fail to effectively mate with the broach with the Calcar Planar Shaft instrument, part number 110032331.
- FDA root cause
- Under Investigation by firm
- Action
- On August 9, 2019 the firm distributed Urgent Medical Device Recall Communication letters by email instructing them to: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to [email protected] within three (3) days. This form must be returned even if you do not have affected products available to return in your territory. b. For each return, send a copy of Attachment 1 to [email protected]. c. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. d. Include a copy of Attachment 2 Certificate of Sterilization with returned instruments. e. Mark RECALL on the outside of the returned cartons. 4. Return the Additional Accounts form to [email protected]. a. Review the list of hospitals included with the email notification sent to your facility, which includes a list of hospitals that have already been notified of this recall. b. Identify whether there are any additional hospitals that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format. c. If there are no additional accounts to notify, please indicate that there are no additional accounts, or indicate None or NA on the form. 5. Retain a copy of your Inventory Return Certification and product return forms for your records in the event of a compliance audit of your facility. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received out
- Quantity in commerce
- 73
- Distribution
- CA, IN, MN, NJ, NY, AR, LA, OH, KY, TX, CO (US) CA, CN, NL (International)
- Product codes and lots
- Model Number: 11032331 Lots: 784060, 662870