Z-0922-2020Class II

Cobas infinity central lab / cobas infinity core license-a Calculator/Data Processing Module, For Clinical...

Roche Diagnostics GmbH / initiated 2019-10-08 / Terminated / root cause: software

Roche Diagnostics Operations, Inc. began a recall of Cobas infinity central lab / cobas infinity core license-a Calculator/Data Processing Module, For Clinical Use Material Number: 07154003001 on . FDA classified it as Class II and gives the reason as "Cobasinfinity laboratory solution Version 2.4.1 through Version 2.5.4 Using a CommServer Driver - Incorrect Alarm Mapping". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0922-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Cobas infinity central lab / cobas infinity core license-a Calculator/Data Processing Module, For Clinical Use Material Number: 07154003001
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Cobasinfinity laboratory solution Version 2.4.1 through Version 2.5.4 Using a CommServer Driver - Incorrect Alarm Mapping
Root cause tag
software
FDA root cause
Software design
Action
Roche issued Urgent Medical Device Correction TP-00791 mailed via UPS Ground (receipt required) on 10/8/2019. The letter identifies the problem, health risk and action to take: Complete the attached fax form (TP-00792) and fax or email it according to the instructions on the form. " File this Urgent Medical Device Correction (UMDC) for future reference Roche will resolve this issue in a future software update. In the next couple of weeks, a Roche Representative will be contacting you to adjust alarm code mapping as needed per your specific configuration. Contact the Roche Support Network Customer Support Center 24 hours a day, seven days a week at 1-800-526-2272 if you have questions. The UMDC will also post to the diagnostics.roche.com website
Quantity in commerce
20
Distribution
US Nationwide in the states of AL, CA, GA, IA, IL, KY, MI, MO, MT, NJ, NY, OR, TN
Product codes and lots
Cobas infinity software version 2.4.1 through 2.5.4

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0922-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0922-2020Class IITerminatedVoluntary: Firm initiated