Cobas infinity central lab / cobas infinity core license-a Calculator/Data Processing Module, For Clinical...
Roche Diagnostics Operations, Inc. began a recall of Cobas infinity central lab / cobas infinity core license-a Calculator/Data Processing Module, For Clinical Use Material Number: 07154003001 on . FDA classified it as Class II and gives the reason as "Cobasinfinity laboratory solution Version 2.4.1 through Version 2.5.4 Using a CommServer Driver - Incorrect Alarm Mapping". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0922-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Cobas infinity central lab / cobas infinity core license-a Calculator/Data Processing Module, For Clinical Use Material Number: 07154003001
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- Cobasinfinity laboratory solution Version 2.4.1 through Version 2.5.4 Using a CommServer Driver - Incorrect Alarm Mapping
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Roche issued Urgent Medical Device Correction TP-00791 mailed via UPS Ground (receipt required) on 10/8/2019. The letter identifies the problem, health risk and action to take: Complete the attached fax form (TP-00792) and fax or email it according to the instructions on the form. " File this Urgent Medical Device Correction (UMDC) for future reference Roche will resolve this issue in a future software update. In the next couple of weeks, a Roche Representative will be contacting you to adjust alarm code mapping as needed per your specific configuration. Contact the Roche Support Network Customer Support Center 24 hours a day, seven days a week at 1-800-526-2272 if you have questions. The UMDC will also post to the diagnostics.roche.com website
- Quantity in commerce
- 20
- Distribution
- US Nationwide in the states of AL, CA, GA, IA, IL, KY, MI, MO, MT, NJ, NY, OR, TN
- Product codes and lots
- Cobas infinity software version 2.4.1 through 2.5.4
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0922-2020 | Class II | Terminated | Voluntary: Firm initiated |