Z-0541-2020Class I
Giraffe Incubator infant warmer
GE Healthcare / initiated 2019-10-11 / Terminated / root cause: design
GE Healthcare, LLC began a recall of Giraffe Incubator infant warmer on . FDA classified it as Class I and gives the reason as "Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0541-2020
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Giraffe Incubator infant warmer
- Product code
- FMT Warmer, Infant Radiant
- Reason for recall
- Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm disseminated its notices by letter beginning on 10/11/2019. The notices identified the following safety issues: - The bedside panels of Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation can be upright and look closed but not be latched. - The portholes also can look closed when not latched. - If a canopy cover is used, it can hold the bedside panel or porthole door closed without being latched. The product may continue to be used following specific safety instructions outlined in the notice. On/about 11/8/2019, the firm issued an "Amended - Urgent Medical Device Recall" letter that supplemented the previous notification to provide additional safety instructions and materials (poster regarding risk of patient fall, user manual addendum, and labeling, including safety labels). Update 6/3/2021: In the event a new consignee was identified, GE Healthcare has sent the original letter.
- Quantity in commerce
- 19425 units
- Distribution
- worldwide
- Product codes and lots
- Model Numbers: 6651-XXXX-XXX 2074921-00X-XXXXXX M1050206 M1050207 M1136120 M1136222 M1136226 M1138367 M1145526 M1145527 M1146091 M1146092 M1148634 M1161351 M1161353 M1161357 M1206149 M1235958 Updated 6/3/2021 - All model numbers and units are affected.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0538-2020 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0539-2020 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0540-2020 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0541-2020 | Class I | Terminated | Voluntary: Firm initiated |