Z-0271-2020Class II
Hitachi Scenaria Whole-body X-ray CT System
FUJIFILM Healthcare Americas Corporation / initiated 2019-10-16 / Terminated / root cause: manufacturing process
Hitachi Medical Systems America Inc began a recall of Hitachi Scenaria Whole-body X-ray CT System on . FDA classified it as Class II and gives the reason as "There is a possibility that the cables to the controller unit may not be affixed properly, creating the potential for a cable to detach during scanning and eject from the system". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0271-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Product
- Hitachi Scenaria Whole-body X-ray CT System
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- There is a possibility that the cables to the controller unit may not be affixed properly, creating the potential for a cable to detach during scanning and eject from the system.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- An Urgent Medical Device Correction Notice dated 10/15/19 was sent to customers. Hitachi will inspect all heat controller units for Scenaria systems S5002-S5014 and S5017-S5044 to ensure that the cables and cable loop clamp are affixed as instructed and apply modifications as needed. We expect to begin this process in late October I early November 2019. This action will be implemented by Hitachi Service at no charge to you. Your Service Area Manager will contact you for scheduling. Please contact me if you need further information regarding this report. You can send any questions, comments or requests to [email protected].
- Quantity in commerce
- 39
- Distribution
- The devices were distributed to the following US states: CA, FL, GA, IA, IN, MD, MT, NC, NY, OH, OK, SD, TN, TX, WY, and PR. The products were distributed to the following foreign countries: Brazil and Mexico.
- Product codes and lots
- Scenaria units S5002-S5014, S5017-S5044
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0271-2020 | Class II | Terminated | Voluntary: Firm initiated |