Z-0271-2020Class II

Hitachi Scenaria Whole-body X-ray CT System

FUJIFILM Healthcare Americas Corporation / initiated 2019-10-16 / Terminated / root cause: manufacturing process

Hitachi Medical Systems America Inc began a recall of Hitachi Scenaria Whole-body X-ray CT System on . FDA classified it as Class II and gives the reason as "There is a possibility that the cables to the controller unit may not be affixed properly, creating the potential for a cable to detach during scanning and eject from the system". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0271-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Hitachi Scenaria Whole-body X-ray CT System
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
There is a possibility that the cables to the controller unit may not be affixed properly, creating the potential for a cable to detach during scanning and eject from the system.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
An Urgent Medical Device Correction Notice dated 10/15/19 was sent to customers. Hitachi will inspect all heat controller units for Scenaria systems S5002-S5014 and S5017-S5044 to ensure that the cables and cable loop clamp are affixed as instructed and apply modifications as needed. We expect to begin this process in late October I early November 2019. This action will be implemented by Hitachi Service at no charge to you. Your Service Area Manager will contact you for scheduling. Please contact me if you need further information regarding this report. You can send any questions, comments or requests to [email protected].
Quantity in commerce
39
Distribution
The devices were distributed to the following US states: CA, FL, GA, IA, IN, MD, MT, NC, NY, OH, OK, SD, TN, TX, WY, and PR. The products were distributed to the following foreign countries: Brazil and Mexico.
Product codes and lots
Scenaria units S5002-S5014, S5017-S5044