Z-0611-2020Class II

EMPOWR Locking Femoral Impactor , Part # 800-05-035, packaged in instrument tray FA EMP PAT TOOL KIT Part #...

Encore Medical, LP / initiated 2019-02-04 / Open, Classified / root cause: design

Encore Medical, Lp began a recall of EMPOWR Locking Femoral Impactor , Part # 800-05-035, packaged in instrument tray FA EMP PAT TOOL KIT Part # 800-99-096 on . FDA classified it as Class II and gives the reason as "The firm has received reports/complaints concerning the breakage of the tips of the locking femoral Impactor. Additionally, if the impactor is not seated correctly on the implant prior to impaction, this may cause off...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0611-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Encore Medical, LP
Product
EMPOWR Locking Femoral Impactor , Part # 800-05-035, packaged in instrument tray FA EMP PAT TOOL KIT Part # 800-99-096
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
The firm has received reports/complaints concerning the breakage of the tips of the locking femoral Impactor. Additionally, if the impactor is not seated correctly on the implant prior to impaction, this may cause off axis loading which can increase the risk of device failure. This could cause a delay in surgery. If the tip is not seen during surgery, long term risks may include an inflammatory response, infection, and/ or pain
Root cause tag
design
FDA root cause
Device Design
Action
On February 4, 2019, DJO emailed consignees a Urgent Field Safety Notice" concerning the recalled product. This field safety notices provided consignees "best practices" for use of the device. On August 19, 2019, DJO revised their "Urgent Field Safety Notice" to communicate the exchange process for physical product in consignees inventory. The field safety notices asked consignees to take the following actions: 1. Please locate within your inventory part 800-05-035 in FA K EMP COMP PAT TOOL KIT, FA K EMP PAT TOOL KIT, and FA K EMP PAT TOOL KIT WIMPCTR. The quantities shown within each consignee letter is specific to the consignee. 2. Contact Customer Service to initiate a return of the product via RMA and to have the replacement impactors sent to you. 3. If you have any questions, please contact Teffany Hutto by phone at (512) 834-6255 or by email at [email protected].
Quantity in commerce
410 devices
Distribution
CO, WA, CA, UT, CA, TN, TX, MS, MO, AR, OK, LA, MN, IL, IN, KY, OH, MI, PA, SC, FL, PR, AL, GA, NM, AZ, NV, VA, RI, MD, ME PA, NY OUS: None
Product codes and lots
All Lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0611-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0611-2020Class IIOngoingVoluntary: Firm initiated