Z-0344-2020Class II
Centurion Medical Products UNIVERSAL LINE FULL BODY DRAPE KIT - REG Code: DT11955
Centurion Medical Products, LP / initiated 2019-10-24 / Terminated / root cause: manufacturing process
Centurion Medical Products Corporation began a recall of Centurion Medical Products UNIVERSAL LINE FULL BODY DRAPE KIT - REG Code: DT11955 on . FDA classified it as Class II and gives the reason as "Sterile packaging incomplete seals may compromise sterility of the product". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0344-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Centurion Medical Products, LP
- Product
- Centurion Medical Products UNIVERSAL LINE FULL BODY DRAPE KIT - REG Code: DT11955
- Product code
- LRO General Surgery Tray
- Reason for recall
- Sterile packaging incomplete seals may compromise sterility of the product
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Centurion medical issued notification was emailed to the affected distributor on October 24, 2019, Centurion's parent company, Med line Industries. Centurion mailed a letter to the single affected medical facility on 10/24/19, via certified mail, return receipt. The letter provides reason for recall, heath risk and action to take: be accessed for use. (1) Identify all inventory that you have within your possession and segregate it in a location where it cannot be accessed for use 2) Please complete the enclosed accountability record, whether or not you have product, and fax or email at your earliest convenience to the attention of Lisa Carpenter as directed. Upon receipt of this record, your response will be documented, and a Centurion Representative will contact you regarding retrieval of the product from your facility, if required. 3) Please forward a copy of this notice to any facilities to whom this product may have been further distributed. Questions Contact Director of Quality Assurance, Centurion Medical Products Corporation (517) 546-5400 Ext.1122
- Quantity in commerce
- 80 units
- Distribution
- Nationwide
- Product codes and lots
- Lot/Batch Number: 2019052190 Exp Date: 5/31/2024
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0344-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY