Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0658-2020Class II

Biopsy Needle FPU Kit, Product # 9731754, UDI: 00643169341517

Medtronic Navigation, Inc / initiated 2019-10-30 / Open, Classified / root cause: manufacturing process

Medtronic Navigation, Inc. began a recall of Biopsy Needle FPU Kit, Product # 9731754, UDI: 00643169341517 on . FDA classified it as Class II and gives the reason as "The firm has become aware that due to a manufacturing issue impacting the biopsy needle depth stop, there exist the potential that some biopsy needle depth stops, when tightened, may not securely tighten to the biopsy...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0658-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic Navigation, Inc
Product
Biopsy Needle FPU Kit, Product # 9731754, UDI: 00643169341517
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
The firm has become aware that due to a manufacturing issue impacting the biopsy needle depth stop, there exist the potential that some biopsy needle depth stops, when tightened, may not securely tighten to the biopsy needle. This could result in a delay in surgery.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 10/24/2019, Medtronic sent a "URGENT: MEDICAL DEVICE RECALL" notification to consignees via FedEx. The customer notification letter as consignees to do the following: 1. Identify, segregate, and quarantine affected products within your inventory 2. Complete the Customer Confirmation Form. Return the form to Medtronic via email at [email protected] or via fax at 651-367-7075 within 30 days of receipt. 3. Contact Medtronic at 1-888-826-5603 to receive a return materials authorization (RMA) and schedule replacement product. Return affected products to: 4. Medtronic Product Services, Attention: RMA #, 1480 Arthur Ave Louisville, CO 80027 5. If you have questions E-mail Medtronic at [email protected] or call 1-888-826-5603.
Quantity in commerce
12 disposable biopsy needles
Distribution
US: Alabama, Alaska,Arizona,Arkansas,California,Colorado,Connecticut,Delaware,District of Columbia,Florida,Georga, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine,Maryland,Massachusetts,Michigan,Minnesota,Mississippi,Missouri,Montana,Nebraska,Nevada,New Hampshire,New Jersey,New Mexico,New York,North Carolina,North Dakota,Ohio,Oklahoma,Oregon,Pennsylvania,South Carolina,South Dakota,Tennessee,Texas,Utah,Vermont,Virginia,Washington,West Virginia,Wisconsin,Wyoming, OUS: None: Australia, Austria,Bangladesh,Belgium,Bosnia And Herzegovina,Brazil,Bulgaria,Canada,Canary Islands,China,Colombia,Costa Rica,Croatia,Czech Republic,Denmark,Ecuador,Egypt,Finland,France,Germany,Greece,Hungary,Iceland,India,Indonesia,Ireland,Israel,Italy,Japan,Jordan,Kazakhstan,Kenya,Korea, Republic Of,Latvia,Lebanon,Luxembourg,Malaysia,Martinique,Mexico,Netherlands,New Zealand,Pakistan,Philippines,Poland,Portugal,Puerto Rico,Qatar,Reunion,Romania,Russian Federation,Saudi Arabia,Serbia,Slovakia,Slovenia,South Africa,Spain,Sweden,Switzerland,Thailand,Turkey,United Arab Emirates,United Kingdom
Product codes and lots
Lot Numbers: 0009665221 0009709452 0009960538

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0658-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0657-2020Class IIOngoingVoluntary: Firm initiated
Z-0658-2020Class IIOngoingVoluntary: Firm initiated
Z-0659-2020Class IIOngoingVoluntary: Firm initiated