Z-0626-2020Class II

C7000 - CUSA Clarity Ultrasonic Tissue Ablation System

Integra LifeSciences Corporation / initiated 2019-10-21 / Terminated / root cause: design

Integra LifeSciences Corp. began a recall of C7000 - CUSA Clarity Ultrasonic Tissue Ablation System on . FDA classified it as Class II and gives the reason as "The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0626-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
C7000 - CUSA Clarity Ultrasonic Tissue Ablation System
Product code
LFL Instrument, Ultrasonic Surgical
Reason for recall
The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.
Root cause tag
design
FDA root cause
Device Design
Action
All impacted consignees (Hospitals, Distributors and Sales reps) were notified by letter delivered by courier service (2nd day), facsimile or email between October 21st and 23rd, 2019. URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION CUSA Clarity Operators Manual Dear Valued Customer/Distributor: The purpose of this letter is to advise you that Integra LifeSciences is voluntarily conducting this field corrective action for the Operators Manuals associated with C7000 CUSA Clarity Console. It was discovered through internal review that we need to revise the Operators Manual to reflect the most accurate sterilization parameters and hence the dry time parameters have been changed from 30 minutes to 40 minutes for the CUSA Clarity 36kHz handpiece and components consistent with the CUSA Clarity 23kHz handpiece and components. There is an improbable likelihood that serious injuries could occur due to the corrections identified within this field action. The internal review confirms that there have been no adverse events or complaints that Integra LifeSciences is aware of associated with the dry cycle time noted in the current Operators Manuals, nor have there been any reports of serious injuries or deaths related to this issue. While it is highly unlikely that a shortened dry time will result in a non-sterile product given that the drying occurs after the steam application, it may result in visible moisture (most likely on the outer wrap). Sterilization Dry Time moisture may be detected by inspection of components and the outer wrap for visible moisture or water. Where visible moisture is noted the product should be re-wrapped and re-sterilized. We request you take the following actions: 1. Customers a. Please review and understand the information provided in Appendix 1. b. Review The Sterilization Parameters in Section 11 of your Operators Manual. c. Replace the information in Section 11 of your Operators Manual with the information
Quantity in commerce
820
Distribution
US: AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS NC NE NJ NY OH OK OR PA SC TN UT VA WA WI WV International: SG MX CA JP AU CO TH HK
Product codes and lots
Swedish OPS Manual  P/N 60905202

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0626-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0613-2020Class IITerminatedVoluntary: Firm initiated
Z-0614-2020Class IITerminatedVoluntary: Firm initiated
Z-0615-2020Class IITerminatedVoluntary: Firm initiated
Z-0616-2020Class IITerminatedVoluntary: Firm initiated
Z-0617-2020Class IITerminatedVoluntary: Firm initiated
Z-0618-2020Class IITerminatedVoluntary: Firm initiated
Z-0619-2020Class IITerminatedVoluntary: Firm initiated
Z-0620-2020Class IITerminatedVoluntary: Firm initiated
Z-0621-2020Class IITerminatedVoluntary: Firm initiated
Z-0622-2020Class IITerminatedVoluntary: Firm initiated
Z-0623-2020Class IITerminatedVoluntary: Firm initiated
Z-0624-2020Class IITerminatedVoluntary: Firm initiated
Z-0625-2020Class IITerminatedVoluntary: Firm initiated
Z-0626-2020Class IITerminatedVoluntary: Firm initiated
Z-0627-2020Class IITerminatedVoluntary: Firm initiated
Z-0628-2020Class IITerminatedVoluntary: Firm initiated
Z-0629-2020Class IITerminatedVoluntary: Firm initiated
Z-0630-2020Class IITerminatedVoluntary: Firm initiated
Z-0631-2020Class IITerminatedVoluntary: Firm initiated
Z-0632-2020Class IITerminatedVoluntary: Firm initiated
Z-0633-2020Class IITerminatedVoluntary: Firm initiated
Z-0634-2020Class IITerminatedVoluntary: Firm initiated
Z-0635-2020Class IITerminatedVoluntary: Firm initiated
Z-0636-2020Class IITerminatedVoluntary: Firm initiated
Z-0637-2020Class IITerminatedVoluntary: Firm initiated
Z-0638-2020Class IITerminatedVoluntary: Firm initiated
Z-0639-2020Class IITerminatedVoluntary: Firm initiated
Z-0640-2020Class IITerminatedVoluntary: Firm initiated
Z-0641-2020Class IITerminatedVoluntary: Firm initiated
Z-0642-2020Class IITerminatedVoluntary: Firm initiated
Z-0643-2020Class IITerminatedVoluntary: Firm initiated
Z-0644-2020Class IITerminatedVoluntary: Firm initiated
Z-0645-2020Class IITerminatedVoluntary: Firm initiated
Z-0646-2020Class IITerminatedVoluntary: Firm initiated
Z-0647-2020Class IITerminatedVoluntary: Firm initiated
Z-0648-2020Class IITerminatedVoluntary: Firm initiated
Z-0649-2020Class IITerminatedVoluntary: Firm initiated
Z-0650-2020Class IITerminatedVoluntary: Firm initiated
Z-0651-2020Class IITerminatedVoluntary: Firm initiated
Z-0652-2020Class IITerminatedVoluntary: Firm initiated
Z-0653-2020Class IITerminatedVoluntary: Firm initiated
Z-0654-2020Class IITerminatedVoluntary: Firm initiated