Biomet Orthopedics Intramedullary Bone Saw Cam Assembly 14MM- Orthopedic manual surgical instrument Item...
Zimmer Biomet, Inc. began a recall of Biomet Orthopedics Intramedullary Bone Saw Cam Assembly 14MM- Orthopedic manual surgical instrument Item Number: 475665 on . FDA classified it as Class II and gives the reason as "Possible commingle of 13 MM and 14 MM Intramedullary Bone Saw Cam Assembly -an Intramedullary Bone Saw Cam Assembly was labeled and etched as a 14 MM but measured as a 13 MM cam". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0595-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Biomet Orthopedics Intramedullary Bone Saw Cam Assembly 14MM- Orthopedic manual surgical instrument Item Number: 475665
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Possible commingle of 13 MM and 14 MM Intramedullary Bone Saw Cam Assembly -an Intramedullary Bone Saw Cam Assembly was labeled and etched as a 14 MM but measured as a 13 MM cam
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Zimmer Biomet issued URGENT MEDICAL DEVICE RECALL letter via FedEx and email on 10/25/19. The letter identifies the reason for recall, health risk and action to take: Review this notification and ensure that affected personnel are aware of the contents. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. Questions call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Alternatively, your questions may be emailed to [email protected].
- Quantity in commerce
- 34 units
- Distribution
- Distribution to US states of CA, FL, IN, MN, MO, NY, and UT, Australia, and New Zealand
- Product codes and lots
- Lot Numbers: 334360, 366680, 366670 UDI: (01)00880304386587(17)280805(10)334360 (01)00880304386587(17)280530(10)366680 (01)00880304386587(17)280530(10)366670
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0594-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0595-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.