Z-0788-2020Class II

Haylard Closed Suction System for Neonates/Pediatrics, 5 Fr, 1. REF 195, Y-Adapter, SAP Finished Product Code...

Halyard Health / initiated 2019-11-05 / Terminated / root cause: manufacturing process

Avanos Medical, Inc. began a recall of Haylard Closed Suction System for Neonates/Pediatrics, 5 Fr, 1. REF 195, Y-Adapter, SAP Finished Product Code 109382303; 2. REF 201, with BALLARD Technology... on . FDA classified it as Class II and gives the reason as "Avanos Medical has received reports stating that the central lumen of some 5 Fr Neonate/Pediatric Closed Suction Catheters may be fully blocked or partially blocked. This may cause the closed suction catheters (CSCs) to...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0788-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Halyard Health
Product
Haylard Closed Suction System for Neonates/Pediatrics, 5 Fr, 1. REF 195, Y-Adapter, SAP Finished Product Code 109382303; 2. REF 201, with BALLARD Technology, SAP Finished Product Code 109382803; 3. REF 202, Elbow, SAP Finished Product Code 109382903; 4. REF 195-5 Y-Adapter, SAP Finished Product Code 109838302.
Product code
BSY Catheters, Suction, Tracheobronchial
Reason for recall
Avanos Medical has received reports stating that the central lumen of some 5 Fr Neonate/Pediatric Closed Suction Catheters may be fully blocked or partially blocked. This may cause the closed suction catheters (CSCs) to inadequately suction secretions and/or saline fluids form the patient's airway to prevent oxygen desaturation.
Root cause tag
manufacturing process
FDA root cause
Manufacturing material removal
Action
Consignees were sent an Urgent Product Advisory Notice letter beginning 11/05/2019 via email and overnight mail. The letter identified affected catalog codes, stated reason for recall, stated that blockage is detectable before catheter use by assessing suctioning effectiveness, and asked for a response form to be returned within 5 business days. questions in the US can be directed to [email protected]. Questions from outside the US can be sent to [email protected].
Quantity in commerce
114,300 units (37,880 US, 76,420 OUS)
Distribution
Nationwide. Internationally distributed to: Argentina, Australia, Austria, Brazil, Canada, Chile, Columbia, Croatia, Czech Republic, Denmark, France, Germany, Greece, Honduras, India, Ireland, Israel, Italy, Korea, Republic of, Lebanon, Mexico, Nepal, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.
Product codes and lots
REF 195 Lot Code / Expiration Date:  M18115T401/2023-APR, M18121T402/2023-MAY, M18128T404/2023-MAY, M18143T401/2023-MAY, M18149T403/2023-MAY, M18156T402/2023-JUN, M18170T402/2023-JUN, M18177T402/2023-JUN, M18192T402/2023-JUL, M18206T401/2023-JUL, M18220T402/2023-AUG, M18247T402/2023-SEP, M18254T405/2023-SEP, M18282T404/2023-OCT, M18311T401/2023-NOV, M18318T402/2023-NOV, M18345T403/2023-DEC, M1834T403/2023-DEC, M18358T403/2023-DEC, M19008T401/2024-JAN, M19058T402/2024-FEB, M19078T403/2024-MAR, M19086T401/2024-MAR, M19098T403/2024-APR, M19134T404/2024-MAY,   REF 201 Lot Code / Expiration Date:  M18044T405/2023-FEB, M18311T402/2023-NOV, M18358T404/2023-DEC, M19037T402/2024-FEB, M19065T402/2024-MAR, M7172T402/2022-JUN.   REF 202 Lot Code / Expiration Date:  M18143T403/2023-MAY, M18192T401/2023-JUL, M18254T404/2023-SEP, M18274T401/2023-OCT, M18296T405/2023-OCT, M18318T401/2023-NOV, M19072T401/2024-MAR, M19142T401/2024-MAY, M7346T403/2022-DEC.   REF 195-5 Lot Code / Expiration Date:  20000095/2024 - JUL, M18274T402/2023-OCT, M18290T402/2023-OCT, M18296T406/2023-OCT, M18311T403/2023-NOV, M18324T401/2023-NOV, M18324T402/2023-NOV, M18332T401/2023-NOV, M18339T401/2023-DEC, M18345T404/2023-DEC, M18352T402/2023-DEC, M19024T402/2024-JAN, M19030T401/2024-JAN, M19051T401/2024-FEB, M19059T401/2024-FEB, M19065T401/2024-MAR, M19072T402/2024-MAR, M19079T401/2024-MAR, M19085T402/2024-MAR, M19098T404/2024-APR, M19099T401/2024-APR, M19105T403/2024-APR, M19120T403/2024-APR, M19128T401/2024-MAY, M19128T403/2024-MAY, M19135T401/2024-MAY, M19142T402/2024-MAY, M19149T405/2024-MAY, M19162T403/2024-JUN, M19177T402/2024-MAY, M19184T401/2024-JUL.