Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0728-2020Class II

S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System. The SMART stent system...

Medtronic, Inc. / initiated 2019-11-14 / Open, Classified / root cause: labeling and UDI

Cardinal Health Inc. began a recall of S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System. The SMART stent system includes a self-expanding stent made of Nitinol... on . FDA classified it as Class II and gives the reason as "Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI Safe"". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0728-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System. The SMART stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) or 7 French (2.3 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diameter when released from the deployment catheter into the iliac artery. Upon deployment, the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2 mm smaller in diameter than the unconstrained diameter of the stent. The SMART Control stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diameter when released from the deployment catheter into the iliac artery. Upon deployment, the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2 mm smaller in diameter than the unconstrained diameter of the stent. The stent is equipped with 12 tantalum markers, 6 at each end, to increase the visibility of the stent under fluoroscopic imaging. The SMART Control stent system is designed to be delivered through an appropriately sized catheter sheath introducer.
Product code
NIO Stent, Iliac
Reason for recall
Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI Safe".
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
Customers were notified with a Medical Device Correction letter. The letter was sent via overnight delivery beginning 11/14/2019. The letter identified affected product, provided a description of the problem, and asked that a copy of the letter be retained where the product is stored. Customer were asked to completed and return an Acknowledgement Form. Questions can be directed to the Cordis Endovascular Medical Information Line 877-DEVICE5 (877-338-4235) Monday through Friday 9:00 am - 5:00 pm EST.
Quantity in commerce
267,247 units total (262,641 OUS, 4,606 US)
Distribution
Worldwide distribution - US Nationwide including Puerto Rico and countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaiian, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Repulic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Maldives, Malta, Martinique, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saint-Barthelemy, Saudia Arabia, Sinapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela.
Product codes and lots
All Lot Codes. Affected SKUs: N1020ABR, N1020BBR, N1020TBR, N1040ABR, N1040BBR, N1040EBR, N1040TBR, N1060ABR, N1060BBR, N1060EBR, N1060TBR, N1080ABR, N1080BBR, N1080EBR, N1080TBR, N6120BV, N620ABR, N620BBR, N620TBR, N640ABR, N640BBR, N640TBR, N660ABR, N660BBR, N660TBR, N680ABR, N680BBR, N680TBR, N7120BV, N720ABR, N720BBR, N720TBR, N740ABR, N740BBR, N740TBR, N760ABR, N760BBR, N760TBR, N780ABR, N780BBR, N780TBR, N8120BV, N820ABR, N820BBR, N820TBR, N840ABR, N840BBR, N840EBR, N840TBR, N860ABR,  N860BBR, N860EBR, N860TBR, N880ABR, N880BBR, N880EBR, N880TBR, N920ABR, N920BBR, N920TBR, N940ABR, N940BBR, N940TBR, N960ABR, N960BBR, N960TBR, N980ABR, N980BBR, N980TBR.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0728-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0728-2020Class IIOngoingVoluntary: Firm initiated
Z-0729-2020Class IIOngoingVoluntary: Firm initiated
Z-0730-2020Class IIOngoingVoluntary: Firm initiated