Proximal Humerus Plate, 8 holes, Right, REF MDS130108R -
Miami Device Solutions, LLC. began a recall of Proximal Humerus Plate, 8 holes, Right, REF MDS130108R - Product Usage: The Proximal Humerus Plating System consists of plates, screws, and locking caps... on . FDA classified it as Class II and gives the reason as "Right Proximal Humerus Plates. Product component parts MDS130108R, may be improperly identified as left". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0603-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Miami Device Solutions, LLC
- Product
- Proximal Humerus Plate, 8 holes, Right, REF MDS130108R - Product Usage: The Proximal Humerus Plating System consists of plates, screws, and locking caps intended to provide internal fixation of fractures of the proximal humerus. The polyaxial screws have a range of motion of +/- 20. The screws can be simply lagged or locked to the plate with the addition of a locking cap. The plates are side specific implants.
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- Right Proximal Humerus Plates. Product component parts MDS130108R, may be improperly identified as left.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm notified their consignees by email beginning on 01/15/2019 requesting the examination of the Proximal Humerus Plates for mismarked units. Return of the units is ongoing.
- Quantity in commerce
- 15 units
- Distribution
- US Nationwide distribution in the states of FL, RI, MA, PA, NY, OH, GA, WA, AL, CA, TX, VA.
- Product codes and lots
- Lot number PM00208
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0603-2020 | Class II | Terminated | Voluntary: Firm initiated |