Z-0683-2020Class II

A1CNow SELF CHECK (European), Software Version Revision D, REF Numbers 3054, 3062

Polymer Technology Systems / initiated 2016-12-02 / Terminated / root cause: software

Polymer Technology Systems began a recall of A1CNow SELF CHECK (European), Software Version Revision D, REF Numbers 3054, 3062 on . FDA classified it as Class II and gives the reason as "A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0683-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Polymer Technology Systems
Product
A1CNow SELF CHECK (European), Software Version Revision D, REF Numbers 3054, 3062
Product code
LCP Assay, Glycosylated Hemoglobin
Reason for recall
A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
Root cause tag
software
FDA root cause
Software design
Action
A Customer Bulletin dated November 2016 was distributed to customers.
Quantity in commerce
500
Distribution
The products were distributed to the following US states: AR, CA, FL, IL, IN, KY, MN, NC, NJ, NY, OH, PA, RI TN, UT, VA, WA, and Puerto Rico. The products were distributed to the following foreign countries: Australia, Bahamas, Belgium, Bahrain, Bangladesh, Canada, China, Columbia, Czech Republic, England, Egypt, Greece, Spain, India, Indonesia, Haiti, Hong Kong, Italy, Japan, Jordan, Kenya, Lebanon, Mexico, Myanmar, Malaysia, Philippines, Pakistan, South Africa, Nigeria, Poland, Saudi Arabia, Sweden, Turkey, Singapore, Trinidad & Tobago, Slovenia, Vietnam, United Arab Emirates, Uruguay.
Product codes and lots
Lot Numbers: 1610622 1618153

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0683-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0682-2020Class IITerminatedVoluntary: Firm initiated
Z-0683-2020Class IITerminatedVoluntary: Firm initiated
Z-0684-2020Class IITerminatedVoluntary: Firm initiated
Z-0685-2020Class IITerminatedVoluntary: Firm initiated
Z-0686-2020Class IITerminatedVoluntary: Firm initiated