HeartStart XL+ Defibrillator/Monitor, Model 861290
Philips North America, LLC began a recall of HeartStart XL+ Defibrillator/Monitor, Model 861290 on . FDA classified it as Class II and gives the reason as "Device may fail to turn on or unexpectedly attempt to restart, rendering it unable to return to a ready for use state. Additionally, this issue may occur when the device is in standby mode, when attempting to power on...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0608-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- HeartStart XL+ Defibrillator/Monitor, Model 861290
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Device may fail to turn on or unexpectedly attempt to restart, rendering it unable to return to a ready for use state. Additionally, this issue may occur when the device is in standby mode, when attempting to power on to run a self-test. If this occurs, the device will indicate that it is not ready for use.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Philips Healthcare issued Medical Device Correction Notices dated 10/3/2019 to customers via certified mail. Customers are advised to take the following actions: - The device is safe to use and can remain in service if it does not exhibit any of these behaviors described in this notice. - If you identify a device that exhibit these behaviors, please remove the device from service and contact Philips to request device. - Fill out, sign and send the reply form that is provided as the last page of the letter.
- Quantity in commerce
- 24,738 units
- Distribution
- US Nationwide
- Product codes and lots
- All Units, Model number 861290