ARCHITECT HAVAB-G Calibrator, REF 6L27-01
Abbott Gmbh & Co. KG began a recall of ARCHITECT HAVAB-G Calibrator, REF 6L27-01 on . FDA classified it as Class II and gives the reason as "Abbott has identified that Calibrator lot 08219BE00 is potentially impacted by an increase in Relative Light Units (RLUs) up to 94%. It has been confirmed that the increase in the calibrator RLU values may cause a...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0574-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- ARCHITECT HAVAB-G Calibrator, REF 6L27-01
- Product code
- LOL Hepatitis A Test (Antibody And Igm Antibody)
- Reason for recall
- Abbott has identified that Calibrator lot 08219BE00 is potentially impacted by an increase in Relative Light Units (RLUs) up to 94%. It has been confirmed that the increase in the calibrator RLU values may cause a decrease in the control values and patient results. Patient values generated using the impacted calibrator lot 08219BE00 may have changed interpretation from reactive to falsely non reactive. Results impacted may present as a (falsely) non reactive result that fall in the range between 0.50 to 0.99 S/CO.
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- The firm initiated the recall by letter on 11/07/2019. The letter explained the issue and requested that the consignee discontinue use of the product and destroy it. The firm is providing replacement material.
- Quantity in commerce
- 83 kits
- Distribution
- US in the states of: AZ, CA, CO, CT, FL, KY, LA, MA, MD, MI, MN, MO, NJ, NY, OH, OR, RI, SC, TN, TX, WA, WV
- Product codes and lots
- lot number 08219BE00
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0574-2020 | Class II | Terminated | Voluntary: Firm initiated |