Z-0718-2020Class II

LIFESTREAM Balloon Expandable Vascular Covered Stent . BARD PERIPHERAL VASCULAR. 6mm X 26mm X 80 Recommended...

Bard Peripheral Vascular Inc / initiated 2019-02-18 / Terminated / root cause: manufacturing process

Bard Peripheral Vascular Inc began a recall of LIFESTREAM Balloon Expandable Vascular Covered Stent . BARD PERIPHERAL VASCULAR. 6mm X 26mm X 80 Recommended Introducer 6F (2.00mm) NP 8atm (811 kPa)... on . FDA classified it as Class II and gives the reason as "Certain product code/lot number combinations may experience introducer sheath compatibility issues when used with a 6F introducer sheath. Using the affected balloon expandable vascular covered stents with a 6F...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0718-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Peripheral Vascular Inc
Product
LIFESTREAM Balloon Expandable Vascular Covered Stent . BARD PERIPHERAL VASCULAR. 6mm X 26mm X 80 Recommended Introducer 6F (2.00mm) NP 8atm (811 kPa) Recommended Guide wire .035" (0.89mm) RBP 12 atm (1216 kPa) RX Only Lot numbers begins with CMA or CMB. Product Code Stent Diameter Stent length Catheter Length LSM0800626 6mm 26 mm 80 cm LSM1350626 6 mm 26 mm 135 cm LSM0800637 6 mm 37 mm 80 cm LSM1350637 6 mm 37 mm 135 cm LSM0800726 7 mm 26 mm 80 cm LSM1350726 7 mm 26 mm 135 cm Product Usage: The LIFESTREAM Balloon Expandable Vascular Covered Stent is indicated for the treatment of atherosclerotic lesions in common and external iliac arteries with reference vessel diameters between 4.5 mm and 12.0 mm, and lesions lengths up to 100mm.
Product code
PRL Iliac Covered Stent, Arterial
Reason for recall
Certain product code/lot number combinations may experience introducer sheath compatibility issues when used with a 6F introducer sheath. Using the affected balloon expandable vascular covered stents with a 6F introducer sheath could result in difficulty or the inability to insert or advance the catheter through the introducer sheath. This could potentially lead to a slight prolongation of the procedure.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 02/18/19, Urgent Product Defect Correction Notices were mailed to customers informing them to upsize to a 7F introducer sheath and to add relabeling stickers over the 6F introducer sheath size indication on the label. In addition, customers were asked to complete and return the Response Form. In addition, if the affected product has been further distributed, immediate action was requested to contact that location, advise them of the field correction notice and to forward the Urgent Product Defect Correction Notice to the affected customer. Customers having additional questions are encouraged to call 1-800-257-232.
Quantity in commerce
127 Units
Distribution
127 units were distributed in Australia to 32 Consignees.
Product codes and lots
Begins with Lot Number CMA or CMB Product Code LSM0800626  Product Code LSM1350626  Product Code LSM0800637  Product Code LSM1350637  Product Code LSM0800726  Product Code LSM1350726

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0718-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0718-2020Class IITerminatedVoluntary: Firm initiated