Z-0672-2020Class II

iTotal CR Knee Replacement System Model Number: TCR-111-1111-010102 -

ConforMIS, Inc. / initiated 2019-10-25 / Terminated / root cause: packaging

Conformis, Inc. began a recall of iTotal CR Knee Replacement System Model Number: TCR-111-1111-010102 - Product Usage: The iTotal CR Knee Replacement System is intended for use as a total knee... on . FDA classified it as Class II and gives the reason as "Medial and lateral polyethylene inserts did not securely fit into the cobalt chrome tray implant". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0672-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ConforMIS, Inc.
Product
iTotal CR Knee Replacement System Model Number: TCR-111-1111-010102 - Product Usage: The iTotal CR Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Medial and lateral polyethylene inserts did not securely fit into the cobalt chrome tray implant.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
Conformis contacted surgeon via phone on 10/16/19 with the replacement details. Replacement inserts that match the implanted tibial tray interlock geometry were shipped to the affected customer.
Quantity in commerce
1 unit
Distribution
US Nationwide distribution in the state of GA.
Product codes and lots
Serial Number: 0450585

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0672-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0672-2020Class IITerminatedVoluntary: Firm initiated