Z-0672-2020Class II
iTotal CR Knee Replacement System Model Number: TCR-111-1111-010102 -
ConforMIS, Inc. / initiated 2019-10-25 / Terminated / root cause: packaging
Conformis, Inc. began a recall of iTotal CR Knee Replacement System Model Number: TCR-111-1111-010102 - Product Usage: The iTotal CR Knee Replacement System is intended for use as a total knee... on . FDA classified it as Class II and gives the reason as "Medial and lateral polyethylene inserts did not securely fit into the cobalt chrome tray implant". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0672-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ConforMIS, Inc.
- Product
- iTotal CR Knee Replacement System Model Number: TCR-111-1111-010102 - Product Usage: The iTotal CR Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Medial and lateral polyethylene inserts did not securely fit into the cobalt chrome tray implant.
- Root cause tag
- packaging
- FDA root cause
- Under Investigation by firm
- Action
- Conformis contacted surgeon via phone on 10/16/19 with the replacement details. Replacement inserts that match the implanted tibial tray interlock geometry were shipped to the affected customer.
- Quantity in commerce
- 1 unit
- Distribution
- US Nationwide distribution in the state of GA.
- Product codes and lots
- Serial Number: 0450585
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0672-2020 | Class II | Terminated | Voluntary: Firm initiated |