Z-0670-2020Class II
Monaco Radiation Treatment Planning System (RTP) System
Elekta Limited / initiated 2019-11-22 / Open, Classified / root cause: software
Elekta Inc began a recall of Monaco Radiation Treatment Planning System (RTP) System on . FDA classified it as Class II and gives the reason as "Monaco is using the incorrect energy when optimizing and calculating dose". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0670-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Elekta Limited
- Product
- Monaco Radiation Treatment Planning System (RTP) System
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- Monaco is using the incorrect energy when optimizing and calculating dose.
- Root cause tag
- software
- FDA root cause
- Software Design Change
- Action
- An Important Field Safety Notification letter is being emailed to affected customers beginning 11/22/2019. The letter identifies affected product, provides a description of the problem, and recommends user not use multiple energies within the same plan until a software fix has been applied to the device. A response form was asked to be returned. Affected customers are advised to contact local Elekta service teams for questions.
- Quantity in commerce
- 76 (7 US, 69 OUS)
- Distribution
- IA, IN, NY, OH, TX, WI. Internationally distributed to: Australia, Canada, China, Denmark, Germany, Greece, Hong Kong, Italy, Netherlands, Sweden, Switzerland, United Kingdom.
- Product codes and lots
- Software Builds: 5.40.00, 5.50.00 and 5.51.00. UDI GTIN: (01)00858164002190(10) 5.40.00, (01)00858164002244(10) 5.50.00, and (01)00858164002268(10) 5.51.00
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0670-2020 | Class II | Ongoing | Voluntary: Firm initiated |