Collagen Reagent, Cat. No. 5368, 2 X 1 mL, For use in platelet aggregation studies, Contains: Collagen...
Helena Laboratories, Corp. began a recall of Collagen Reagent, Cat. No. 5368, 2 X 1 mL, For use in platelet aggregation studies, Contains: Collagen (equine tendon) For In-Vitro Diagnostic Use on . FDA classified it as Class III and gives the reason as "Through an investigation, it has been determined that vials of Collagen packaged within the Collagen Reagent and Platelet aggregation Kit, may contain particulate matter or microbial growth". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0537-2020
- Classification
- Class III
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Helena Laboratories, Corp.
- Product
- Collagen Reagent, Cat. No. 5368, 2 X 1 mL, For use in platelet aggregation studies, Contains: Collagen (equine tendon) For In-Vitro Diagnostic Use,
- Product code
- GHR Reagent, Platelet Aggregation
- Reason for recall
- Through an investigation, it has been determined that vials of Collagen packaged within the Collagen Reagent and Platelet aggregation Kit, may contain particulate matter or microbial growth.
- Root cause tag
- packaging
- FDA root cause
- Unknown/Undetermined by firm
- Action
- On 05/23/2018, the firm's technical services contacted consignees via phone call with a email follow-up of their recall notification letter. Consignees were advised of the following: 1. Prevent any shipment or schedule demonstrations with these lots 2. Remove the lots from available inventory and discard COLLAGEN vials. 3. Verify amount REMOVED FROM STOCK 4. COMPLETE THE FORM AND RETURN TO HELENA LABORATORIES, USA VIA: fax: 409-842-1874 OR E-MAIL TO [email protected]. For questions, contact Jessica W. Jones, MD, MPH - Mgr., Technical Services at: 1 (409) 842-3714 ext. 1177
- Quantity in commerce
- 64 Aggregation Kits
- Distribution
- US:TX, WV, OH, TN, GA, TN, NY, ME, MA,LA, AZ OUS: Israel, Korea, Australia, Indonesia, Canada, Thailand, Colombia, Brasil , China , Taiwan
- Product codes and lots
- Component Lot # 5-17-5368 (Collagen), Exp. 5/31/19 Kit Lot # 3-17-5369
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0536-2020 | Class III | Ongoing | Voluntary: Firm initiated | ||
| Z-0537-2020 | Class III | Ongoing | Voluntary: Firm initiated |