Z-1439-2020Class II

bellavista 1000 ventilator, Catalog number 301.100.030. bellavista is an electronically controlled pneumatic...

Jaeger Medical GmbH / initiated 2019-11-25 / Terminated / root cause: software

Vyaire Medical began a recall of bellavista 1000 ventilator, Catalog number 301.100.030. bellavista is an electronically controlled pneumatic ventilation system with an integrated air... on . FDA classified it as Class II and gives the reason as "The G6 bellavista 1000 US ventilators may experience intermittent failures: Lack of acoustic high priority alarm, presence of a 'no alarm' condition, or presence of non-responsive touch screen". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1439-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Jaeger Medical GmbH
Product
bellavista 1000 ventilator, Catalog number 301.100.030. bellavista is an electronically controlled pneumatic ventilation system with an integrated air compressing system. bellavista uses room air and high-pressure oxygen. Air enters through a fresh gas intake port and is compressed together with the oxygen by the blower. Oxygen enters through a high pressure inlet. An electronic mixer valve provides for the operator-set concentration. Gas is supplied to the patient via the microprocessor controlled inspiratory valve.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
The G6 bellavista 1000 US ventilators may experience intermittent failures: Lack of acoustic high priority alarm, presence of a 'no alarm' condition, or presence of non-responsive touch screen.
Root cause tag
software
FDA root cause
Software design
Action
Consignees were sent an Urgent Medical Device Recall letter beginning 11/25/2019 via overnight delivery. The letter identified affected devices, provided a description of the issue, and provided actions to be taken by distributors and end-users. A response form was asked to be returned. Questions can be directed to Vyaire's Technical Support Department at [email protected].
Quantity in commerce
116
Distribution
US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, MD, MI, PA, TX, WA and Puerto Rico.
Product codes and lots
Serial numbers: MB170104, MB170118, MB170151, MB170202, MB230000, MB230020, MB230024, MB230027, MB230031, MB230033, MB230035, MB230036, MB230037, MB230038, MB230039, MB230040, MB230041, MB230046, MB230048, MB230049, MB230052, MB230053, MB230054, MB230055, MB230056, MB230057, MB230058, MB230059, MB230060, MB230061, MB230062, MB230063, MB230064, MB230065, MB230066, MB230067, MB230068, MB230069, MB230070, MB230071, MB230072, MB230073, MB230074, MB230075, MB230076, MB230077, MB230078, MB230079, MB230080, MB230081, MB230082, MB230083, MB230084, MB230085, MB230086, MB230087, MB230088, MB230089, MB230090, MB230091, MB230092, MB230093, MB230094, MB230095, MB230097, MB230098, MB230099, MB230100, MB230101, MB230102, MB230104, MB230190, MB230191, MB230194, MB230195, MB230196, MB230197, MB230201, MB230204, MB230205, MB230207, MB230209, MB230210, MB230217, MB230219, MB230220, MB230221, MB230222, MB230227, MB230228, MB230229, MB230230, MB230231, MB230232, MB230233, MB230243, MB230244, MB230246, MB230247, MB230249, MB230251, MB230253, MB230256, MB230258, MB230260, MB230261, MB230262, MB230263, MB230266, MB230270, MB230274, MB230275, MB230277, MB230294, MB230297, MB230298.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1439-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1439-2020Class IITerminatedVoluntary: Firm initiated