Z-0773-2020Class II
Invia Foam Dressing Kit with FitPad - X-Large (3 pcs) SKU # 0876224. The Invia Foam Dressing Kit with FitPad...
Medela AG / initiated 2019-12-02 / Terminated / root cause: packaging
Medela Inc began a recall of Invia Foam Dressing Kit with FitPad - X-Large (3 pcs) SKU # 0876224. The Invia Foam Dressing Kit with FitPad consists of a foam pad, a section interface (Invia... on . FDA classified it as Class II and gives the reason as "Invia Foam Dressing Kit with FitPad could have holes in the sterile packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0773-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medela AG
- Product
- Invia Foam Dressing Kit with FitPad - X-Large (3 pcs) SKU # 0876224. The Invia Foam Dressing Kit with FitPad consists of a foam pad, a section interface (Invia FitPad) and Transparent Film (one or more pieces depending on the kit size).
- Product code
- OMP Negative Pressure Wound Therapy Powered Suction Pump
- Reason for recall
- Invia Foam Dressing Kit with FitPad could have holes in the sterile packaging.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- An Important Device Recall letter was sent to customers via certified mail beginning 12/02/2019. The letter identified affected product, stated the reason for recall, and customers can return or destroy affected product. A response form was asked to be returned. Questions can be addressed by email to [email protected] or call 888-835-5968.
- Quantity in commerce
- 331
- Distribution
- AL, AZ, CA, CO, FL, GA, IL, MN, MO, NC, NY, PA, SC, TN, TX, WA
- Product codes and lots
- Lots: 17110012731 / Exp 17-Nov-19, 17110022736 / Exp 22-Nov-19, 17120012746 / Exp 8-Dec-19, 18010012767 / 4-Jan-20, 18070012913 / 31-Jul-20, 18090012960 / Exp 28-Sep-20, 18110013046 / 5-Nov-20.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0773-2020 | Class II | Terminated | Voluntary: Firm initiated |