3M Durapore Advanced, Surgical Tape, Model Numbers: 1590-0, 1590-1, 1590-2, 1590-3, 1591-0, 1591-1, 1591-2...
3M Company - Health Care Business began a recall of 3M Durapore Advanced, Surgical Tape, Model Numbers: 1590-0, 1590-1, 1590-2, 1590-3, 1591-0, 1591-1, 1591-2, 1591-3, 1590S-1, 1590S-2, 1590B-118, 1590B-136 on . FDA classified it as Class II and gives the reason as "While the product meets all safety and efficacy criteria, long term exposure of 3M Durapore Advanced Surgical Tape to continuous high moisture clinical applications has not fully met all of 3M s performance...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0719-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- 3M Company
- Product
- 3M Durapore Advanced, Surgical Tape, Model Numbers: 1590-0, 1590-1, 1590-2, 1590-3, 1591-0, 1591-1, 1591-2, 1591-3, 1590S-1, 1590S-2, 1590B-118, 1590B-136
- Product code
- KGX Tape And Bandage, Adhesive
- Reason for recall
- While the product meets all safety and efficacy criteria, long term exposure of 3M Durapore Advanced Surgical Tape to continuous high moisture clinical applications has not fully met all of 3M s performance expectations. This use environment may cause tape bond failure, leading the tape to not perform as intended. Further, the tape fibers may fray which could be accidentally ingested or aspirated.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm initiated the recall by letter on 11/20/2019. The letter explained the failures seen and requested the following actions be taken: 1. Immediately cease use of the product and quarantine it for destruction/disposal. 2. Complete the attached Customer Acknowledgement Form with the following: - Acknowledge that you have read and understood this letter whether or not you have any 3M" Durapore" Advanced Surgical Tape in stock at your facility; and - Indicate if you have any product in stock and dispose of this material per facility policy. - Email the completed form to [email protected] 3. If there are clinical questions, contact Kimberly Schommer, RN, BSN via the 3M Health Care Customer Helpline at 1-800-228-3957.
- Quantity in commerce
- 470,954 rolls
- Distribution
- Nationwide, Canada, Germany, France, Switzerland, Denmark