Paltop Dynamic Dental Implant Dia 3.25mm L 13.0 mm, Catalog Number 21-70020
Paltop Advanced Dental Solutions Ltd began a recall of Paltop Dynamic Dental Implant Dia 3.25mm L 13.0 mm, Catalog Number 21-70020 on . FDA classified it as Class II and gives the reason as "Some of the implants were packaged with a Standard Platform Cover Screw instead of a Narrow Platform Cover Screw. The Standard Platform Cover Screw will not fit in the Narrow Platform implant". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0836-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Paltop Advanced Dental Solutions Ltd
- Product
- Paltop Dynamic Dental Implant Dia 3.25mm L 13.0 mm, Catalog Number 21-70020
- Product code
- DZE Implant, Endosseous, Root-Form
- Reason for recall
- Some of the implants were packaged with a Standard Platform Cover Screw instead of a Narrow Platform Cover Screw. The Standard Platform Cover Screw will not fit in the Narrow Platform implant.
- Root cause tag
- packaging
- FDA root cause
- Mixed-up of materials/components
- Action
- Urgent: Medical Device Recall notification letters dated 11/18/19 were sent to customers. Please discontinue use. Attached is a Reply Communication form that should be used to let us know whether the product has been used or will be returned. Once we receive that Form, a Customer Service agent will reach out to you with return instructions and provide a label if necessary. Paltop Advanced Dental will replace the implants that are in stock.
- Quantity in commerce
- 67
- Distribution
- The products were distributed to the following US states: CA, DC, FL, GA, MA, MD, NJ, NY, SC, TX, and VA.
- Product codes and lots
- UDI: 07290108695649; Lot Number WO-011320