Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0836-2020Class II

Paltop Dynamic Dental Implant Dia 3.25mm L 13.0 mm, Catalog Number 21-70020

Paltop Advanced Dental Solutions Ltd / initiated 2019-11-18 / Terminated / root cause: packaging

Paltop Advanced Dental Solutions Ltd began a recall of Paltop Dynamic Dental Implant Dia 3.25mm L 13.0 mm, Catalog Number 21-70020 on . FDA classified it as Class II and gives the reason as "Some of the implants were packaged with a Standard Platform Cover Screw instead of a Narrow Platform Cover Screw. The Standard Platform Cover Screw will not fit in the Narrow Platform implant". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0836-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Paltop Advanced Dental Solutions Ltd
Product
Paltop Dynamic Dental Implant Dia 3.25mm L 13.0 mm, Catalog Number 21-70020
Product code
DZE Implant, Endosseous, Root-Form
Reason for recall
Some of the implants were packaged with a Standard Platform Cover Screw instead of a Narrow Platform Cover Screw. The Standard Platform Cover Screw will not fit in the Narrow Platform implant.
Root cause tag
packaging
FDA root cause
Mixed-up of materials/components
Action
Urgent: Medical Device Recall notification letters dated 11/18/19 were sent to customers. Please discontinue use. Attached is a Reply Communication form that should be used to let us know whether the product has been used or will be returned. Once we receive that Form, a Customer Service agent will reach out to you with return instructions and provide a label if necessary. Paltop Advanced Dental will replace the implants that are in stock.
Quantity in commerce
67
Distribution
The products were distributed to the following US states: CA, DC, FL, GA, MA, MD, NJ, NY, SC, TX, and VA.
Product codes and lots
UDI: 07290108695649; Lot Number WO-011320

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0836-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0836-2020Class IITerminatedVoluntary: Firm initiated