Z-0780-2020Class II

Stryker T2 Alpha Delta Strike Plate IMN Instruments - part of the IMN Instruments System and is used to...

Stryker Corp. / initiated 2019-12-05 / Terminated

Stryker GmbH began a recall of Stryker T2 Alpha Delta Strike Plate IMN Instruments - part of the IMN Instruments System and is used to support insertion and withdrawal of nails through the... on . FDA classified it as Class II and gives the reason as "Instrument can break at the level of the thread when being exposed to high forces during nail implantation or removal which can lead to complications during nail insertion and removal". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0780-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Stryker T2 Alpha Delta Strike Plate IMN Instruments - part of the IMN Instruments System and is used to support insertion and withdrawal of nails through the Targeting Device Catalog Number: 2351-0050
Product code
HRS Plate, Fixation, Bone
Reason for recall
Instrument can break at the level of the thread when being exposed to high forces during nail implantation or removal which can lead to complications during nail insertion and removal.
FDA root cause
Under Investigation by firm
Action
Stryker issued Urgent Medical Device Recall Letter (PFA# 2219868) to US consignees on 5-Dec-2019 , letter states reason for recall, health risk and action to take: 1.Please inform users of this Urgent Medical Device Recall and forward this notice to all those individuals who need to be aware within your organization. 2.Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. A response is required, even though you may not have any physical inventory on site anymore. 3.Quarantine and discontinue use of the recalled devices. 4. Hospitals/Branches/Agencies: Complete and sign the enclosed Urgent Medical Device Recall Business Reply. An OUS customer letter was also issued at the country level.
Quantity in commerce
433 (289 units U.S. / 144 units OUS)
Distribution
Worldwide distribution - US Nationwide distribution and countries of Switzerland, France, Spain, United Kingdom, Italy, Sweden, Netherlands, Australia.
Product codes and lots
All lots UDI: (01)07613327350210

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0780-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0780-2020Class IITerminatedVoluntary: Firm initiated