Stryker T2 Alpha Delta Strike Plate IMN Instruments - part of the IMN Instruments System and is used to...
Stryker GmbH began a recall of Stryker T2 Alpha Delta Strike Plate IMN Instruments - part of the IMN Instruments System and is used to support insertion and withdrawal of nails through the... on . FDA classified it as Class II and gives the reason as "Instrument can break at the level of the thread when being exposed to high forces during nail implantation or removal which can lead to complications during nail insertion and removal". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0780-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stryker T2 Alpha Delta Strike Plate IMN Instruments - part of the IMN Instruments System and is used to support insertion and withdrawal of nails through the Targeting Device Catalog Number: 2351-0050
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- Instrument can break at the level of the thread when being exposed to high forces during nail implantation or removal which can lead to complications during nail insertion and removal.
- FDA root cause
- Under Investigation by firm
- Action
- Stryker issued Urgent Medical Device Recall Letter (PFA# 2219868) to US consignees on 5-Dec-2019 , letter states reason for recall, health risk and action to take: 1.Please inform users of this Urgent Medical Device Recall and forward this notice to all those individuals who need to be aware within your organization. 2.Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. A response is required, even though you may not have any physical inventory on site anymore. 3.Quarantine and discontinue use of the recalled devices. 4. Hospitals/Branches/Agencies: Complete and sign the enclosed Urgent Medical Device Recall Business Reply. An OUS customer letter was also issued at the country level.
- Quantity in commerce
- 433 (289 units U.S. / 144 units OUS)
- Distribution
- Worldwide distribution - US Nationwide distribution and countries of Switzerland, France, Spain, United Kingdom, Italy, Sweden, Netherlands, Australia.
- Product codes and lots
- All lots UDI: (01)07613327350210
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0780-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
