Fast-Cath Hemostasis Introducer, HEMOSTASIS INTRODUCER, 8.5F, 2.83 mm, REF 406143 -
Abbott Medical began a recall of Fast-Cath Hemostasis Introducer, HEMOSTASIS INTRODUCER, 8.5F, 2.83 mm, REF 406143 - Product Usage: The Fast-Cath Hemostasis Introducers are indicated for the... on . FDA classified it as Class II and gives the reason as "Three lots of Fast-Cath Hemostasis Introducer devices contained 8F sized dilators and sheaths instead of the expected 8.5F devices". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0789-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Fast-Cath Hemostasis Introducer, HEMOSTASIS INTRODUCER, 8.5F, 2.83 mm, REF 406143 - Product Usage: The Fast-Cath Hemostasis Introducers are indicated for the introduction of angiographic catheters, closed-end catheters, balloon catheters, and electrodes into a vessel where minimizing blood loss is essential.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Three lots of Fast-Cath Hemostasis Introducer devices contained 8F sized dilators and sheaths instead of the expected 8.5F devices.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- Beginning on December 9, 2019, US consignees will be notified in person by an Abbott field representative. OUS consignees will be notified by mail. The firm is seeking return of the devices.
- Quantity in commerce
- 380 devices
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of MA, VA, NH, FL, CT, UT, OH, WI, NY, IA, MI, MS, PA, and countries of Canada, Czech Republic
- Product codes and lots
- Batch 6917183; UDI (01)15414734014219(17)220228(10)6917183 Batch 6950628; UDI (01)15414734014219(17)220331(10)6950628 Batch 7099144; UDI (01)15414734014219(17)220630(10)7099144
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0789-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.