Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0789-2020Class II

Fast-Cath Hemostasis Introducer, HEMOSTASIS INTRODUCER, 8.5F, 2.83 mm, REF 406143 -

Abbott Medical / initiated 2019-12-09 / Terminated / root cause: packaging

Abbott Medical began a recall of Fast-Cath Hemostasis Introducer, HEMOSTASIS INTRODUCER, 8.5F, 2.83 mm, REF 406143 - Product Usage: The Fast-Cath Hemostasis Introducers are indicated for the... on . FDA classified it as Class II and gives the reason as "Three lots of Fast-Cath Hemostasis Introducer devices contained 8F sized dilators and sheaths instead of the expected 8.5F devices". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0789-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Fast-Cath Hemostasis Introducer, HEMOSTASIS INTRODUCER, 8.5F, 2.83 mm, REF 406143 - Product Usage: The Fast-Cath Hemostasis Introducers are indicated for the introduction of angiographic catheters, closed-end catheters, balloon catheters, and electrodes into a vessel where minimizing blood loss is essential.
Product code
DYB Introducer, Catheter
Reason for recall
Three lots of Fast-Cath Hemostasis Introducer devices contained 8F sized dilators and sheaths instead of the expected 8.5F devices.
Root cause tag
packaging
FDA root cause
Packaging
Action
Beginning on December 9, 2019, US consignees will be notified in person by an Abbott field representative. OUS consignees will be notified by mail. The firm is seeking return of the devices.
Quantity in commerce
380 devices
Distribution
Worldwide distribution - US Nationwide distribution in the states of MA, VA, NH, FL, CT, UT, OH, WI, NY, IA, MI, MS, PA, and countries of Canada, Czech Republic
Product codes and lots
Batch 6917183; UDI (01)15414734014219(17)220228(10)6917183  Batch 6950628; UDI (01)15414734014219(17)220331(10)6950628  Batch 7099144; UDI (01)15414734014219(17)220630(10)7099144

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0789-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0789-2020Class IITerminatedVoluntary: Firm initiated