BIOMERIEUX VIDAS 3, Compact immunoanalyzer with full traceability and automation
Biomerieux Inc began a recall of BIOMERIEUX VIDAS 3, Compact immunoanalyzer with full traceability and automation on . FDA classified it as Class II and gives the reason as "Following Customers complaints, investigations have been initiated on potentially false results obtained on VIDAS 3 with an expired calibration. Indeed, the calibrations of assays were valid in the calibration menu...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0928-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- BIOMERIEUX VIDAS 3, Compact immunoanalyzer with full traceability and automation
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Following Customers complaints, investigations have been initiated on potentially false results obtained on VIDAS 3 with an expired calibration. Indeed, the calibrations of assays were valid in the calibration menu whereas in fact the calibrations were expired and no alarm displayed to warn the users about the expiration date of the calibrations. Software computes results of analysis for assays with expired calibrations. The anomaly is due to an incorrect update of the calibration status by the software.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm initiated their field action by an email to their subsidiaries on 10/22/2019. The subsidiaries then notified their customers by 2-day express via Fed Ex. In the US customer letter, customers are instructed to check their software version. If the VIDAS 3 software version is between 1.2.0 and 1.3.1, customers are instructed to check the archiving mode. If the archiving mode is Automatic When Results are Reported then customers are instructed how to configure the workaround, and to contact bioMerieux if the customer has the incorrect software configuration. bioMerieux then instructs impacted customers to review and interpret results in the context of the overall clinical procedure. Once this is completed to complete the acknowledgment form in Attachment A and return it the local bioMerieux representative and confirm receipt of this notice.
- Quantity in commerce
- 738 units
- Distribution
- US
- Product codes and lots
- software versions 1.2 and higher
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0928-2020 | Class II | Terminated | Voluntary: Firm initiated |