Z-0920-2020Class II

BA-800M Chemistry Analyzer Part Numbers BA81E-PA00002 and BA81E-PA00003, Containing Gas spring...

Datascope Corp. / initiated 2019-12-09 / Terminated / root cause: manufacturing process

Mindray DS USA, Inc. dba Mindray North America began a recall of BA-800M Chemistry Analyzer Part Numbers BA81E-PA00002 and BA81E-PA00003, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring))... on . FDA classified it as Class II and gives the reason as "The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0920-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
BA-800M Chemistry Analyzer Part Numbers BA81E-PA00002 and BA81E-PA00003, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring)) and Gas spring YQ6/15-84-235(B-B)-80N (033-000090-00 (rear spring))
Product code
JGS Electrode, Ion Specific, Sodium
Reason for recall
The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The consignee notification was sent certified mail with return receipt verification. The letter identified the affected product, problem and actions to be taken by the Customer/User. Customers were instructed to: Continue to use the system normally and exercise caution when lifting/closing the front and rear cover. Attachment A includes a list of system(s) identified by model and serial number purchased by your company. We recommend that you notify the clients to which you have distributed these systems and advise them to continue normal use with the recommended caution. The gas springs on the systems listed in Attachment A will require replacement and instructions are provided in Attachment B. The parts needed to perform the replacement will be shipped to you within the next several weeks. Type of Action by the Company Mindray will provide replacement gas springs at no cost. Should you require technical assistance while preforming the replacement, please contact Mindrays Technical Support team at 800-288-2121, option 2 (Technical Support), then choose option 3 (IVD Support team).
Quantity in commerce
7
Distribution
The products were distributed to the following US states: MI and TX.
Product codes and lots
Serial Numbers: W1-6C000006,  W1-71000007,  W1-73000009,  W1-73000010,  XP-68000687T,  W1-71000008,  W1-85000011

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0920-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0920-2020Class IITerminatedVoluntary: Firm initiated
Z-0921-2020Class IITerminatedVoluntary: Firm initiated