Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0809-2020Class II

Medline Thermistor Foley Catheter 14 FR 5-10 ml, non sterile Catalog Number: 102201101463MD {Medline code...

Degania Silicone, Inc. / initiated 2019-12-02 / Terminated / root cause: manufacturing process

Degania Silicone, Ltd. began a recall of Medline Thermistor Foley Catheter 14 FR 5-10 ml, non sterile Catalog Number: 102201101463MD {Medline code 55346) Monitor urinary output and bladder temperature... on . FDA classified it as Class II and gives the reason as "Temperature deviation did not meet release specifications". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0809-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Degania Silicone, Inc.
Product
Medline Thermistor Foley Catheter 14 FR 5-10 ml, non sterile Catalog Number: 102201101463MD {Medline code 55346) Monitor urinary output and bladder temperature in addition to facilitating urine drainage.
Product code
EZL Catheter, Retention Type, Balloon
Reason for recall
Temperature deviation did not meet release specifications
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, Q Medical Degania issued an "URGENT FIELD SAFETY NOTICE" letter dated 02-DEC-2019 via email, the letter states reason for recall, health risk and action to take: Please forward this information to the immediate attention of the persons responsible for usage of the affected parts. Sub-recall must be initiated by your company to collect any distributed product from the suspected list. 2. Please fill in the attached Recall Response form and indicate the quantity of the affected product in your possession. 3. Please promptly return the completed Recall Response form to Degania Silicone by e-mail: [email protected]. 4. The disposition of the recalled goods should be scrap. We would ask you to destruct the goods at your facility and send us the Scrap confirmation letter. Should you have any questions about this subject, please contact Degania Silicone: Degonia Silicone ltd. Degania Bet, 1513000, ISRAEL Katia Kayam, Customer complaints Officer, Tel: +972-529532059 E-mail: [email protected]
Quantity in commerce
2080 units
Distribution
US Distribution to IL
Product codes and lots
Lot Number: V18046701

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0809-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0809-2020Class IITerminatedVoluntary: Firm initiated