Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0837-2020Class II

Arrow Endurance" Extended Dwell Peripheral Catheter System, REF: ASK-00820-FLH

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2019-12-06 / Terminated / root cause: manufacturing process

Arrow International Inc began a recall of Arrow Endurance" Extended Dwell Peripheral Catheter System, REF: ASK-00820-FLH on . FDA classified it as Class II and gives the reason as "Potential sterility issue for specific lots of the Arrow Endurance" Extended Dwell Peripheral Catheter System, Product Code ASK-00820-FLH". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0837-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
Arrow Endurance" Extended Dwell Peripheral Catheter System, REF: ASK-00820-FLH
Product code
FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Reason for recall
Potential sterility issue for specific lots of the Arrow Endurance" Extended Dwell Peripheral Catheter System, Product Code ASK-00820-FLH.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
On December 6, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail stating: Urgent Medical Device Recall Notice Arrow Endurance" Extended Dwell Peripheral Catheter System EIF-000391 December 06, 2019 To: Customer of Arrow Products Arrow International is voluntarily recalling the following product code and lots: Product Code: ASK-00820-FLH Lot Numbers: 13F19E0247, 13F19F0006, 13F19G0433 Arrow International is recalling the above product code and lots due to a potential sterility issue. The product was processed through the incorrect sterilization cycle; thus, the sterility of the device cannot be guaranteed. If the devices sterility is compromised and the product is used it could lead to infectious complications. No complaints, patient injuries or deaths have been reported at this time. Our records indicate you have received products that are subject to this action. We are now notifying our customers to take the following actions: 1. Inspect your inventory for the affected product codes above. If you have affected stock in inventory, immediately discontinue use and quarantine any products with the product codes and lot numbers listed above. 2. To acknowledge receipt of this Urgent Medical Device Recall Notice or to return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document your receipt of this letter. Adverse reactions or quality problems experienced with the us
Quantity in commerce
575
Distribution
US: FL
Product codes and lots
REF: ASK-00820-FLH Manufacturing Lots: 13F19E0247, 13F19F0006, and 13F19G0433 Exp. Date: Aug. 2020 - Sep. 2020

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0837-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0837-2020Class IITerminatedVoluntary: Firm initiated