Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0996-2020Class II

GORE EXCLUDER Iliac Branch Endoprosthesis (IBE)

W. L. Gore & Associates, Inc. / initiated 2020-01-06 / Terminated / root cause: manufacturing process

W. L. Gore & Associates Inc. began a recall of GORE EXCLUDER Iliac Branch Endoprosthesis (IBE) on . FDA classified it as Class II and gives the reason as "Firm has received reports of leading end catheter component separations". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0996-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
W. L. Gore & Associates, Inc.
Product
GORE EXCLUDER Iliac Branch Endoprosthesis (IBE)
Product code
MIH System, Endovascular Graft, Aortic Aneurysm Treatment
Reason for recall
Firm has received reports of leading end catheter component separations.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The recall consists of corrections with no proposed product removal. The correction will consist of: A Medical Device Safety Correction letter sent to US physicians and hospitals and OUS hospitals and Distributors, along with an IFU Summary of Changes document Updates to the products Instruction for Use Updates to physician didactic training Additional manufacturing process error-proofing The firm is also instructing customers to share the letter with others in their hospital or clinic as appropriate, and contact the firm's Customer Service (email: [email protected] or by phone at 800-528- 8763) with any questions related to the letter. Any adverse event should be reported to the manufacturer and the country specific regulatory authorities immediately. To report an event to Recalling Firm, email: [email protected] or contact: USA: +1.800.528.1866, Ext. 44922, +1.928.864.4922, Fax: +1.928.864.4364 China: +86 21 5172 8237, Fax: +86 21 5172 8236 Japan: +81 3 6746 2562, Fax: +81 3 6746 2563 Brazil: +55 11 5502-7953, Fax: +55 11 5502-7965 EMEA: +49 89 4612 3440, Fax: +49 89 4612 43440
Quantity in commerce
23212 units
Distribution
US: OUS: Australia; Austria; Canada; China; Taiwan; United Arab Emirates; Argentina; Brazil; Switzerland; Chile; Colombia; Costa Rica; Czechia; Spain; Greece; Croatia; Hungary; Ireland; Israel; Italy; United Kingdom; Lebanon; Mexico; New Zealand; Panama; Poland; Saudi Arabia; Slovenia; Germany; Netherlands; Norway; Finland; France; Sweden; Monaco;
Product codes and lots
-Catalog numbers PLC*, PLA*, PLL*, RLT23*, RLT26*, RLT28*, RLT31*, RLT35*, HGB*, and CEB* *= size  -not specific to any particular lot or serial numbers.  For full details of affected products see 806 Attachment 1 - Catalog Numbers and Sizes.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0996-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0995-2020Class IITerminatedVoluntary: Firm initiated
Z-0996-2020Class IITerminatedVoluntary: Firm initiated