LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting...
Smith & Nephew, Inc. began a recall of LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting... on . FDA classified it as Class II and gives the reason as "Multiple lots of LEGION AP Femoral Cutting Blocks were manufactured with an oversize green plunger hole and therefore do not meet their design specifications". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0878-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Smith & Nephew, Inc.
- Product
- LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 71434413. (Orthopaedic manual surgical instrument).
- Product code
- MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
- Reason for recall
- Multiple lots of LEGION AP Femoral Cutting Blocks were manufactured with an oversize green plunger hole and therefore do not meet their design specifications.
- Root cause tag
- manufacturing process
- FDA root cause
- Process change control
- Action
- The firm initiated the recall by letter on December 16, 2019. The letter identified the problem and directed the consignee to remove the affected lots noted in the Urgent Medical Device Recall Notice.
- Quantity in commerce
- 59 units
- Distribution
- USA: IA, NC, PA, TN. Int'l: CH, BE, GB, AE, CN, DE, NL, BR, FR.
- Product codes and lots
- LEGION AP Cutting Block Size 4, Product Number 71434409, Lot Numbers: 17LM05399, 17LM07852, 17LM07852A; LEGION AP Cutting Block Size 6, Product Number 71434411, Lot Numbers: 17LM05397, 17LM05398, 17LM07849; LEGION AP Cutting Block Size 7, Product Number 71434412, Lot Numbers: 17LM01746, 17LM05390, 17LM05394, 17LM07851; LEGION AP Cutting Block Size 8, Product Number 71434413, Lot Number: 17LM08442.