Brasseler USA KM169-39-76 Steinmann Pin Single Diamond Smooth, Diameter 2.8 mm, Length 228 mm
Brasseler USA I Lp began a recall of Brasseler USA KM169-39-76 Steinmann Pin Single Diamond Smooth, Diameter 2.8 mm, Length 228 mm Product Usage: A Steinmann Pin is used for internal fixation of... on . FDA classified it as Class II and gives the reason as "103 packaged and labeled tubes (103 product tubes with 6 pins per tube) of KM168-39-76 Single Diamond Smooth Steinmann Pins were labeled with the incorrect product labels. The tubes were incorrectly labeled as...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0884-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Brasseler USA I, L.P.
- Product
- Brasseler USA KM169-39-76 Steinmann Pin Single Diamond Smooth, Diameter 2.8 mm, Length 228 mm Product Usage: A Steinmann Pin is used for internal fixation of large bone fractures.
- Product code
- HTY Pin, Fixation, Smooth
- Reason for recall
- 103 packaged and labeled tubes (103 product tubes with 6 pins per tube) of KM168-39-76 Single Diamond Smooth Steinmann Pins were labeled with the incorrect product labels. The tubes were incorrectly labeled as KM169-39-76 Single Diamond Threaded Steinmann Pins.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Initial contact to affected consignees was made on 12/11/2019 by telephone to provide notification of the labeling error and arrange for replacement product. Recall letter also sent to affected consignees.
- Quantity in commerce
- 56 packs of 6
- Distribution
- US: AZ, AR, FL
- Product codes and lots
- Lot Number NR8ZR