Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0884-2020Class II

Brasseler USA KM169-39-76 Steinmann Pin Single Diamond Smooth, Diameter 2.8 mm, Length 228 mm

Brasseler USA I, L.P. / initiated 2019-12-11 / Terminated / root cause: labeling and UDI

Brasseler USA I Lp began a recall of Brasseler USA KM169-39-76 Steinmann Pin Single Diamond Smooth, Diameter 2.8 mm, Length 228 mm Product Usage: A Steinmann Pin is used for internal fixation of... on . FDA classified it as Class II and gives the reason as "103 packaged and labeled tubes (103 product tubes with 6 pins per tube) of KM168-39-76 Single Diamond Smooth Steinmann Pins were labeled with the incorrect product labels. The tubes were incorrectly labeled as...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0884-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Brasseler USA I, L.P.
Product
Brasseler USA KM169-39-76 Steinmann Pin Single Diamond Smooth, Diameter 2.8 mm, Length 228 mm Product Usage: A Steinmann Pin is used for internal fixation of large bone fractures.
Product code
HTY Pin, Fixation, Smooth
Reason for recall
103 packaged and labeled tubes (103 product tubes with 6 pins per tube) of KM168-39-76 Single Diamond Smooth Steinmann Pins were labeled with the incorrect product labels. The tubes were incorrectly labeled as KM169-39-76 Single Diamond Threaded Steinmann Pins.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Initial contact to affected consignees was made on 12/11/2019 by telephone to provide notification of the labeling error and arrange for replacement product. Recall letter also sent to affected consignees.
Quantity in commerce
56 packs of 6
Distribution
US: AZ, AR, FL
Product codes and lots
Lot Number NR8ZR

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0884-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0884-2020Class IITerminatedVoluntary: Firm initiated