Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources...
GE Healthcare, LLC began a recall of Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources, and from other healthcare information sources on . FDA classified it as Class II and gives the reason as "Centricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0819-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources, and from other healthcare information sources.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Centricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm initiated their correction on 12/16/2019 by letter. The letter explained the issue and directed the consignee to follow instructions in the letter until an update to the Centricity Universal Viewer can be made by the firm to correct the issue.
- Quantity in commerce
- 205
- Distribution
- Distributed nationwide; Internationally distributed to Mexico and 24 additional countries.
- Product codes and lots
- Model 2088026-9XX Versions 6.0 SP10 or higher
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0819-2020 | Class II | Terminated | Voluntary: Firm initiated |