Medonic M-Series Hematology Analyzer M16S BD ABR US
Clinical Diagnostic Solutions, Inc. began a recall of Medonic M-Series Hematology Analyzer M16S BD ABR US Product code: 1400075; OUS Product code 1400011 on . FDA classified it as Class II and gives the reason as "A mix-up of autosampler tube positions with a possibility of misidentification of sample results and a risk of operator injury from an exposed aspiration needle". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1001-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Clinical Diagnostic Solutions, Inc.
- Product
- Medonic M-Series Hematology Analyzer M16S BD ABR US Product code: 1400075; OUS Product code 1400011
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- A mix-up of autosampler tube positions with a possibility of misidentification of sample results and a risk of operator injury from an exposed aspiration needle.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Consignees notified via email and postal delivery with recall letter dated 01/02/2020. The letter identified affected product, stated reason for recall, and provided specific actions to follow. A Recall Acknowledgement Form was requested to be returned by email, postal delivery, or fax. Clinical Diagnostic Solutions Inc. is handling US consignees only. Boule Medical AB is handling OUS consignees.
- Quantity in commerce
- 261
- Distribution
- Worldwide distribution. US Nationwide, Venezuela, Africa, Italy, Brazil, and Ukraine. No governmental, Canadian, or Mexican consignees.
- Product codes and lots
- Part Number 1400075 (US) Part Number 1400011 (OUS) All serial numbers distributed prior to 3/21/2019
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1001-2020 | Class II | Terminated | Voluntary: Firm initiated |