Z-1001-2020Class II

Medonic M-Series Hematology Analyzer M16S BD ABR US

Clinical Diagnostic Solutions, Inc. / initiated 2020-01-02 / Terminated / root cause: software

Clinical Diagnostic Solutions, Inc. began a recall of Medonic M-Series Hematology Analyzer M16S BD ABR US Product code: 1400075; OUS Product code 1400011 on . FDA classified it as Class II and gives the reason as "A mix-up of autosampler tube positions with a possibility of misidentification of sample results and a risk of operator injury from an exposed aspiration needle". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1001-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Clinical Diagnostic Solutions, Inc.
Product
Medonic M-Series Hematology Analyzer M16S BD ABR US Product code: 1400075; OUS Product code 1400011
Product code
GKZ Counter, Differential Cell
Reason for recall
A mix-up of autosampler tube positions with a possibility of misidentification of sample results and a risk of operator injury from an exposed aspiration needle.
Root cause tag
software
FDA root cause
Software design
Action
Consignees notified via email and postal delivery with recall letter dated 01/02/2020. The letter identified affected product, stated reason for recall, and provided specific actions to follow. A Recall Acknowledgement Form was requested to be returned by email, postal delivery, or fax. Clinical Diagnostic Solutions Inc. is handling US consignees only. Boule Medical AB is handling OUS consignees.
Quantity in commerce
261
Distribution
Worldwide distribution. US Nationwide, Venezuela, Africa, Italy, Brazil, and Ukraine. No governmental, Canadian, or Mexican consignees.
Product codes and lots
Part Number 1400075 (US) Part Number 1400011 (OUS) All serial numbers distributed prior to 3/21/2019

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1001-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1001-2020Class IITerminatedVoluntary: Firm initiated