Z-0941-2020Class II
Safeline Y-Type Blood Set, Catalog Number V2500
B Braun Medical Inc / initiated 2019-11-20 / Terminated / root cause: manufacturing process
B. Braun Medical, Inc. began a recall of Safeline Y-Type Blood Set, Catalog Number V2500 on . FDA classified it as Class II and gives the reason as "Potential for leakage at the joint between the blood filters and tubing". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0941-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- B Braun Medical Inc
- Product
- Safeline Y-Type Blood Set, Catalog Number V2500
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- Potential for leakage at the joint between the blood filters and tubing
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 11/20/2019, B Braun Medical Inc. issued Urgent Medical Device letters and Product Removal Acknowledgment forms to customers. On 11/20/2019, B Braun Medical Inc. also issued a press release. Actions Required By BBMI Customer/User: 1. Review the Device Recall Notification in its entirety and ensure that all users in your organization of the mentioned product, and other concerned persons, are informed about this voluntary product recall. If you are a distributor, please forward a copy of the recall notification to your customers. 2. Determine your current inventory of the affected lots within your facility, cease use and quarantine product subject to recall. Do not destroy any affected product. 3. Utilizing the Product Removal Acknowledgement, form, record the total number of individual units (within partial cases) and the number of full-unopened cases. If you have no inventory remaining, please enter zero (0) on the form. 4. Return the completed Product Removal Acknowledgement form to B. Braun Medical Inc. Quality Assurance department by faxing the form to (610) 849-1197 or e-mail to PA_QualityAssurance. 5. If you have any full cases, partial cases or unused individual pieces of these affected products as identified in the Product Removal Acknowledgement form that was submitted to BBMI Quality Assurance Department, a BBMI Customer Support Representative will contact you to provide instructions for handling the affected product and arrange for return to BBMI. This may include credit or replacement product. Should you identify product where this issue has occurred, please report the event promptly to BBMI by contacting our Clinical and Technical Support Department at 1-800-854-6851.
- Quantity in commerce
- 19,500 units
- Distribution
- US Nationwide
- Product codes and lots
- GUIDID: 04046964188633 Lot Numbers: 61674884 - Expiration Date: April 30, 2024, 61677634 - Expiration Date: April 30, 2024, 61678262 - Expiration Date: May 31, 2024, 61680625 - Expiration Date: May 31, 2024, 61685110 - Expiration Date: June 30, 2024
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0935-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0936-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0937-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0938-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0939-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0940-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0941-2020 | Class II | Terminated | Voluntary: Firm initiated |