Vaginal Verification Panel Ref 8208 Lot 8208-11
Microbiologics Inc began a recall of Vaginal Verification Panel Ref 8208 Lot 8208-11 on . FDA classified it as Class II and gives the reason as "Product 8208 Vaginal Verification Panel is a 6 pool kit and has contaminant Candida krusei in Pool 2 (subcomponent 6063). Candida krusei is an organism intended to be contained in Pool 6 of the kit. This should have no...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1096-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Microbiologics, Inc.
- Product
- Vaginal Verification Panel Ref 8208 Lot 8208-11
- Product code
- OHQ Multi-Analyte Controls Unassayed
- Reason for recall
- Product 8208 Vaginal Verification Panel is a 6 pool kit and has contaminant Candida krusei in Pool 2 (subcomponent 6063). Candida krusei is an organism intended to be contained in Pool 6 of the kit. This should have no impact on patient test results as long as Gardnerella vaginalis was properly detected when using Pool 2 and/or users are strictly using this product for instrument verification/validation.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Letters will be sent via email to all consignees and end users regarding the correction. Users are to review the instructions, determine how the information affects their usage, and upon determination either use with new information or discard the product.
- Quantity in commerce
- 9
- Distribution
- Distributed in the US to 7 states: CA, KS, ME, NC, NY, OR, and UT. No Canada, Mexico or VA.
- Product codes and lots
- Lot 8208-11
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1096-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY