Z-1150-2020Class II
ROSA Brain 3.0 Application-Brain
MedTech SAS / initiated 2020-01-06 / Terminated / root cause: software
MEDTECH SAS began a recall of ROSA Brain 3.0 Application-Brain on . FDA classified it as Class II and gives the reason as "Some cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1150-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- MedTech SAS
- Product
- ROSA Brain 3.0 Application-Brain
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Some cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the surgery planning.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On January 6, 2020, Zimmer Biomet issued URGENT MEDICAL DEVICE CORRECTION notices to customers via FedEx. Actions Required By Users: Your device can continue to be used, as long as the recommendations below are followed: While the current Instructions For Use do not require the images to be reviewed with ROSA Brain software prior to surgery, they adequately describe the standard workflow and indicate that the image acquisition of the patients head (MRI / CT or 2D X-Ray images) is performed prior to surgery, and is then loaded into the device and used by the surgeon in the preoperative phase to carry out the surgical planning. Due to this potential image management issue, Zimmer Biomet recommends that images be reviewed using the ROSA Brain software prior to surgery. If an issue is detected with the reconstruction and display of the images using the ROSA software, and if the artifacts are located in the Region Of Interest and prevent the surgical planning, then the surgery might have to be cancelled or converted to traditional surgery. However, if the artifacts are not located inside the Region Of Interest and if the surgery planning is successful, the affected device can continue to be used. Transmission of the Medical Device Correction: Please advise the appropriate personnel working in your department with the ROSA system of the content of the letter. Risk Manager Responsibilities: 1. Review the notification and ensure affected personnel are aware of the contents. 2. Complete the Certificate of Acknowledgement. a. Return a digital copy to [email protected]. b. Retain a copy of the Acknowledgement Form with your field action records in the event of a compliance audit of your facilities documentation. 3. If you have further questions or concerns after reviewing the notice, please call Perry Twyford at 281-389-3236 between 9:00 am and 6:00pm EST, Monday through Friday. Alternatively, your questions may be e
- Quantity in commerce
- 47 units
- Distribution
- US Nationwide
- Product codes and lots
- Serial Numbers: BR16004, BR16005, BR16006, BR16009, BR16010, BR16011, BR16012, BR16014, BR16015, BR16016, BR16017, BR16018, BR16021, BR16023, BR16025, BS16904, BS16905, BS16910, BS16916, BS16917, BS16923, Additional Serial Numbers / UDI: 1) BR17030 / UDI - (01)03760244031754(11)180309(21)BR17030; 2) BR17034, BS17934 / UDI - (01)03760244031754(11)180315(21)BR17034; 3) BR17035, BS17935 / UDI - (01)03760244031754(11)180315(21)BR17035; 4) BR17036 / UDI - (01)03760244031754(11)180417(21)BR17036; 5) BR17037, BS17937 / UDI - (01)03760244031754(11)180313(21)BR17037; 6) BR18024 / UDI - (01)03760244031754(11)180321(21)BR18024; 7) BR18031 / UDI - (01)03760244031754(11)180309(21)BR18031; 8) BR18038 / UDI - BS18938 "(01)03760244031754(11)180430(21)BR18038; 9) BR18039 / UDI - (01)03760244031754(11)180416(21)BR18039; 10) BR18040, BS18940 / UDI - (01)03760244031754(11)180427(21)BR18040; 11) BR18041, BS18941 / UDI - (01)03760244031754(11)180329(21)BR18041; 12) BR18042 / UDI - (01)03760244031754(11)180420(21)BR18042; 13) BR18043, BS18943 / UDI - (01)03760244031754(11)180503(21)BR18043; 14) BR18045, BS18945 / UDI - (01)03760244031754(11)180504(21)BR18045; 15) BR18046, BS18946 / UDI - (01)03760244031754(11)180515(21)BR18046; 16) BR18047 / UDI - (01)03760244031754(11)180413(21)BR18047; 17) BR18048, BS18948 / UDI - (01)03760244031754(11)180608(21)BR18048; 18) BR18049, BS18949 / UDI - (01)03760244031754(11)180625(21)BR18049; 19) BR18050 / UDI - (01)03760244031754(11)180531(21)BR18050; 20) BR18051, BS18951 / UDI - (01)03760244031754(11)180605(21)BR18051; 21) BR18052, BS18952 / UDI - (01)03760244031754(11)180621(21)BR18052; 22) BR18053 / UDI - (01)03760244031754(11)180718(21)BR18053; 23) BR18054 / UDI - (01)03760244031754(11)180710(21)BR18054; 24) BR18055 / UDI - (01)03760244031754(11)180605(21)BR18055; 25) BR18056, BS18956 / UDI - (01)03760244031754(11)180611(21)BR18056; 26) BR18057 / UDI - (01)03760244031754(11)180614(21)BR18057; 27) BR18058 / UDI - (01)03760244031754(11)180717(21)BR18058; 28) BR18059, BS18959 - UDI - (01)03760244031754(11)180627(21)BR18059; 29) BR18060 / UDI - (01)03760244031754(11)180627(21)BR18060; 30) BR18061, BS18961 / UDI - (01)03760244031754(11)180704(21)BR18061; 31) BR18062 / UDI - (01)03760244031754(11)180806(21)BR18062; 32) BR18063 / UDI - (01)03760244031754(11)180807(21)BR18063; 33) BR18064 / UDI - (01)03760244031754(11)180806(21)BR18064; 34) BR18065 / UDI - (01)03760244031754(11)180823(21)BR18065; 35) BR18066, BS18966 / UDI - (01)03760244031754(11)180827(21)BR18066; 36) BR18067 / UDI - (01)03760244031754(11)180903(21)BR18067; 37) BR18068, BS18968 / UDI - (01)03760244031754(11)180813(21)BR18068; 38) BR18069, BS18969 / UDI - (01)03760244031754(11)180820(21)BR18069; 39) BR18070, BS18970 / UDI - (01)03760244031754(11)180829(21)BR18070; 40) BR18071, BS18971 / UDI - (01)03760244031754(11)180904(21)BR18071; 41) BR18072, BS18972 / UDI - (01)03760244031754(11)180831(21)BR18072; 42) BR18073 / UDI - (01)03760244031754(11)180910(21)BR18073; 43) BR18074, BS18974 / UDI - (01)03760244031754(11)181108(21)BR18074; 44) BR18075, BS18975 / UDI - (01)03760244031754(11)180906(21)BR18075; 45) BR18076, BS18976 / UDI - (01)03760244031754(11)180907(21)BR18076; 46) BR18077, BS18977 / UDI - (01)03760244031754(11)180911(21)BR18077; 47) BR18078 / UDI - (01)03760244031754(11)180912(21)BR18078; 48) BR18079 / UDI - (01)03760244031754(11)180912(21)BR18079; 49) BR18080 / UDI - (01)03760244031754(11)180913(21)BR18080; 50) BR18081 / UDI - (01)03760244031754(11)180926(21)BR18081; 51) BR18082 / UDI - (01)03760244031754(11)181009(21)BR18082; 52) BR18083 / UDI - (01)03760244031754(11)181011(21)BR18083; 53) BR18084, BS18984 / UDI - (01)03760244031754(11)181108(21)BR18084; 54) BR18085 / UDI - (01)03760244031754(11)181017(21)BR18085; 55) BR18086, BS18986 / UDI - (01)03760244031754(11)181019(21)BR18086; 56) BR18087, BS18987 / UDI - (01)03760244031754(11)181112(21)BR18087; 57) BR18088, BS18988 / UDI - (01)03760244031754(11)181114(21)BR18088; 58) BR18089 / UDI - (01)03760244031754(11)181109(21)BR18089; 59) BR18090 / UDI - (01)03760244031754(11)181120(21)BR18090; 60) BR18091 / UDI - (01)03760244031754(11)181112(21)BR18091; 61) BR18092 / UDI - (01)03760244031754(11)181128(21)BR18092; 62) BR18093 / UDI - (01)03760244031754(11)181116(21)BR18093; 63) BR18094 / UDI - (01)03760244031754(11)181119(21)BR18094; 64) BR18095, BS18995 / UDI - (01)03760244031754(11)181121(21)BR18095; 65) BR18096, BS18996 / UDI - (01)03760244031754(11)181126(21)BR18096; 66) BR18097 / UDI - (01)03760244031754(11)181129(21)BR18097; 67) BR18098, BS18998 / UDI - (01)03760244031754(11)181123(21)BR18098; 68) BR18099 / UDI - (01)03760244031754(11)181211(21)BR18099; 69) BR18100 / UDI - (01)03760244031754(11)181217(21)BR18100; 70) BR18101 / UDI - (01)03760244031754(11)181221(21)13R18101
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1150-2020 | Class II | Terminated | Voluntary: Firm initiated | ||