Z-1151-2020Class II

ROSA One 3.1 Brain Application

MedTech SAS / initiated 2020-01-06 / Terminated / root cause: software

MEDTECH SAS began a recall of ROSA One 3.1 Brain Application on . FDA classified it as Class II and gives the reason as "Some cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1151-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
MedTech SAS
Product
ROSA One 3.1 Brain Application
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Some cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the surgery planning.
Root cause tag
software
FDA root cause
Software design
Action
On January 6, 2020, Zimmer Biomet issued URGENT MEDICAL DEVICE CORRECTION notices to customers via FedEx. Actions Required By Users: Your device can continue to be used, as long as the recommendations below are followed: While the current Instructions For Use do not require the images to be reviewed with ROSA Brain software prior to surgery, they adequately describe the standard workflow and indicate that the image acquisition of the patients head (MRI / CT or 2D X-Ray images) is performed prior to surgery, and is then loaded into the device and used by the surgeon in the preoperative phase to carry out the surgical planning. Due to this potential image management issue, Zimmer Biomet recommends that images be reviewed using the ROSA Brain software prior to surgery. If an issue is detected with the reconstruction and display of the images using the ROSA software, and if the artifacts are located in the Region Of Interest and prevent the surgical planning, then the surgery might have to be cancelled or converted to traditional surgery. However, if the artifacts are not located inside the Region Of Interest and if the surgery planning is successful, the affected device can continue to be used. Transmission of the Medical Device Correction: Please advise the appropriate personnel working in your department with the ROSA system of the content of the letter. Risk Manager Responsibilities: 1. Review the notification and ensure affected personnel are aware of the contents. 2. Complete the Certificate of Acknowledgement. a. Return a digital copy to [email protected]. b. Retain a copy of the Acknowledgement Form with your field action records in the event of a compliance audit of your facilities documentation. 3. If you have further questions or concerns after reviewing the notice, please call Perry Twyford at 281-389-3236 between 9:00 am and 6:00pm EST, Monday through Friday. Alternatively, your questions may be e
Quantity in commerce
39 units
Distribution
US Nationwide
Product codes and lots
Serial Number / UDI Number:  1) BS19037 / UDI - (01)03760244030948(11)190423(21)BS19037;  2) BS19038 / UDI - (01)03760244030948(11)190415(21)BS19038;  3) BS19039 / UDI - (01)03760244030948(11)190430(21)BS19039; 4) BS19040 / UDI - (01)03760244030948(11)190513(21)BS19040; 5) BS19041 / UDI - (01)03760244030948(11)190523(21)BS19041; 6) BS19043 / UDI - (01)03760244030948(11)190513(21)BS19043;  7) BS19049 / UDI - (01)03760244032492(11)190730(21)BS19049;  8) BS19050 / UDI - (01)03760244032492(11)190726(21)BS19050; 9) BS19052 / UDI - (01)03760244032492(11)190731(21)BS19052;  10) BS19054 / UDI - (01)03760244032492(11)190912(21)BS19054;  11) BS19057 / UDI - (01)03760244032492(11)191003(21)BS19057;  12) BS19058 / UDI - (01)03760244032492(11)191011(21)BS19058.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1151-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1150-2020Class IITerminatedVoluntary: Firm initiated
Z-1151-2020Class IITerminatedVoluntary: Firm initiated