SIGNA Architect Nuclear Magnetic Resonance Imaging System -
GE Healthcare, LLC began a recall of SIGNA Architect Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high... on . FDA classified it as Class II and gives the reason as "It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0875-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- SIGNA Architect Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- This issue will be resolved by sending an Urgent Medical Device Correction letter to Medical Facility staff (Director of Clinical/Radiology, Risk Manager/Hospital Administrator, Director of Biomedical Engineering) with affected systems. This letter will inform the customer of the issue and instruct the user to ensure the displayed system date is correct. The long-term solution is to correct affected systems with updated software.
- Quantity in commerce
- 62 (30 US, 32 OUS) total devices
- Distribution
- Worldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
- Product codes and lots
- Mfg. Lot or Serial # System ID PG75A1700023MR 630DMGMR3T 33-4-1252 312702MR3T 033-4-1005 518262MR750W To be provided 082427210212 To be provided 082427170026 To be provided GON4713311 To be provided M4017073 To be provided EM0601 To be provided O009MR03 To be provided GON4634845
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0874-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0875-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0876-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0877-2020 | Class II | Terminated | Voluntary: Firm initiated |