Z-0917-2020Class II

Bard Conquest 40 PTA Dilation Catheter, 8mm X 20mm x 75cm (CQF7582), 8mm x 40mm x 75cm (CQF7584); Recommended...

Bard Peripheral Vascular Inc / initiated 2019-08-14 / Terminated / root cause: labeling and UDI

Bard Peripheral Vascular Inc began a recall of Bard Conquest 40 PTA Dilation Catheter, 8mm X 20mm x 75cm (CQF7582), 8mm x 40mm x 75cm (CQF7584); Recommended Guidewire .035"; Recommended Introducer 6F... on . FDA classified it as Class II and gives the reason as "The catheter packaging incorrectly listed the balloon size as 8mm X 20mm instead of the intended size 8mm X 40mm. If undetected, the user will advance the longer than expected balloon to the lesion. The potential exists...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0917-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Peripheral Vascular Inc
Product
Bard Conquest 40 PTA Dilation Catheter, 8mm X 20mm x 75cm (CQF7582), 8mm x 40mm x 75cm (CQF7584); Recommended Guidewire .035"; Recommended Introducer 6F; Nominal Pressure 8atm, Rated Burst Pressure 40atm; Syringe inflation 1cc; UDI: (01)00801741060533(17)220228(10)REDQ3444
Product code
LIT Catheter, Angioplasty, Peripheral, Transluminal
Reason for recall
The catheter packaging incorrectly listed the balloon size as 8mm X 20mm instead of the intended size 8mm X 40mm. If undetected, the user will advance the longer than expected balloon to the lesion. The potential exists that upon inflation, non-diseased vessel, kidney or sections of the AV dialysis fistulae may be dilated, potentially causing a varying degree of injury. This may require additional medical or surgical intervention.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
On 8/14/2019, a "Urgent Medical Device Recall Notification was sent to consignees via FedEx. Shortly after sending the original customer letter on 08/14/2019, the firm realized that customers may experience issues identifying the affected Lot Number in their inventory if they searched by the Product Code CQF7584. On 9/10/2019, the firm sent a 2nd notice explaining that although CQF7584 is the affected Product Code that they purchased, the actual Product Code listed on the package labeling is CQF7582. In addition to providing details about the recall products, the notification ask customer to do the following: 1. Our records show that your facility has purchased the product code affected by this voluntary recall. Do not use or further distribute any affected product. 2. Please check all inventory locations within your institution for the affected product code / lot number combination listed in the recall notice. If you have further distributed this product code / lot number, please immediately contact that location, advise them of the recall, forward these instructions and have them return the affected product to BPV. 3. Please remove any identified product from your shelves. 4. If you have used the affected product, complete and return the attached Recall and Effectiveness Check Form indicating no product will be returned. Once the product affected by this recall has been removed from your inventory: 5. Fill out the Recall and Effectiveness Check Form. Be sure to state the quantities and lot numbers of each recalled product that you intend to return. It is extremely important that we receive this information. 6. Email the completed Recall and Effectiveness Check Form to [email protected] or fax it to BPV at 1-800-994-6772. 7. Once the Recall and Effectiveness Check Form has been completed and emailed to [email protected] or faxed to BPV at 1-800-994-6772 and all information has been verified, the BPV C
Quantity in commerce
286 Units
Distribution
US: NJ, NY, OH, PA, VA, SC, GA, AR, TX, MA, NC, TN, LA, AZ, MS, FL, AL,MD, CA, HI, MO, CO, WI, DC, KY, IL, MN, IN, KS, OK, CT, MT, WA, ND, IA, WA, WV, PR OUS: None
Product codes and lots
Lot Number REDQ3444 Exp: 02/28/2022

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0917-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0917-2020Class IITerminatedVoluntary: Firm initiated