Siemens ARTIS pheno, Model # 10849000. Interventional Fluoroscopic X-Ray System -
Siemens Medical Solutions USA, Inc began a recall of Siemens ARTIS pheno, Model # 10849000. Interventional Fluoroscopic X-Ray System - Product Usage: Angiography system for single and biplane diagnostic imaging... on . FDA classified it as Class II and gives the reason as "The potential exists for system movement to be permanently blocked by an activation of the collision sensor integrated in the cover of the collimator even when no collision has occurred". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0807-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens ARTIS pheno, Model # 10849000. Interventional Fluoroscopic X-Ray System - Product Usage: Angiography system for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- The potential exists for system movement to be permanently blocked by an activation of the collision sensor integrated in the cover of the collimator even when no collision has occurred.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Siemens Medical Solutions USA, Inc. sent a Customer Safety Advisory Notice to all affected customers via AX070/19/S on 12/20/2019.
- Quantity in commerce
- 6
- Distribution
- US Nationwide distribution in the states of FL, IL, IN, MA, PA, VA.
- Product codes and lots
- Serial Numbers 164231, 164234, 164240, 164241, 164242, 164244
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0807-2020 | Class II | Terminated | Voluntary: Firm initiated |