Z-0853-2020Class II
VitreQ 27G VFI Cannula-REF: CN27.D03 GTIN: 8719214221461
Vitreq BV / initiated 2019-12-17 / Terminated / root cause: packaging
Vitreq Bv began a recall of VitreQ 27G VFI Cannula-REF: CN27.D03 GTIN: 8719214221461 on . FDA classified it as Class II and gives the reason as "Microscopic tears of the sterile pouch may compromise sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0853-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Vitreq BV
- Product
- VitreQ 27G VFI Cannula-REF: CN27.D03 GTIN: 8719214221461
- Product code
- HMX Cannula, Ophthalmic
- Reason for recall
- Microscopic tears of the sterile pouch may compromise sterility
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- VitreQ issued Urgent Medical Device recall dated 12/16/19 letter states reason for recall, health risk and action to take: 1.Immediately examine your inventory and quarantine product from all lots subject to recall. If you have further distributed this product, please identify your customers and notify them at once of this product recall. Consider all potential users of this product in your user supply chain. You may provide a copy of this recall notification letter; 2.Complete the Attachment 1: Response Form enclosed IMMEDIATELY, as evidence of the product being returned, and we will credit your account or complete the response form even if you do not have product to return.3.Return the Attachment 1: Response Form to BVI by one of these methods: Fax 866-906-4304 or email a copy to [email protected]. If you have any specific product replacement questions you may contact Customer Service Team at: [email protected] or call 866-906-8080 from 8 A.M. to 6 P.M. EST (Monday-Friday).
- Distribution
- Nationwide
- Product codes and lots
- All lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0846-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0847-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0848-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0849-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0850-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0851-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0852-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0853-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0854-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0855-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0856-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0857-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0858-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0859-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0860-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0861-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.