Terumo Advanced Perfusion System 1 Roller Pump, 4 Inch, Catalog Number 816570 -
Terumo Cardiovascular Systems Corporation began a recall of Terumo Advanced Perfusion System 1 Roller Pump, 4 Inch, Catalog Number 816570 - Product Usage: is indicated for use for up to 6 hours in the extracorporeal... on . FDA classified it as Class II and gives the reason as "These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1108-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Terumo Advanced Perfusion System 1 Roller Pump, 4 Inch, Catalog Number 816570 - Product Usage: is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical proffessional who is experienced in the operation of this or similar equipment. The roller pump is used to move fluids through the cardiopulmonary bypass circuit and can be used for a number of applicatinos including arterial blood pumping, cardioplegia delivery, suction, and venting.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Terumo began notifying customers via telephone to schedule the repairs on 1/2/20. Urgent Medical Device Correction notification letters dated 1/6/20 were sent to customers.
- Quantity in commerce
- 33
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CT, FL, LA, MA, MI, NY, and TX. Countries of Hong Kong, India, South Korea, and Taiwan.
- Product codes and lots
- UDI: 00886799000533; Serial Numbers: 6860 6861 6862 6863 6864 6865 6876 6877 6885 6910 6911 6912 6913 6934 6935 6936 6937 6938 6939 6940 6941 6942 6943 6944 6945 6946 6947 6948 6949 6950 6951 6952 6953
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1108-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1109-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1110-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1111-2020 | Class II | Terminated | Voluntary: Firm initiated |