Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0865-2020Class II

Medline DYNDA1928 VP TRAY WITH PVP Packaged in Mexico GLOVE:Made in Malaysia APPLICATOR:Made in China STERILE...

MEDLINE INDUSTRIES, LP - Northfield / initiated 2020-01-07 / Terminated / root cause: packaging

Medline Industries Inc began a recall of Medline DYNDA1928 VP TRAY WITH PVP Packaged in Mexico GLOVE:Made in Malaysia APPLICATOR:Made in China STERILE (01)40889942587583(10)0000000000 (17)000000(30)20... on . FDA classified it as Class II and gives the reason as "There is potential for the sterile wrap to become loose while in transit and compromise the sterile barrier". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0865-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline DYNDA1928 VP TRAY WITH PVP Packaged in Mexico GLOVE:Made in Malaysia APPLICATOR:Made in China STERILE (01)40889942587583(10)0000000000 (17)000000(30)20 Packaged for Medline Industries, Inc., Northfield, IL 60093 USA. 1-800-MEDLINE www.medline.com WINGED SPONGE:Made in Mexico
Product code
LRP Tray, Surgical
Reason for recall
There is potential for the sterile wrap to become loose while in transit and compromise the sterile barrier.
Root cause tag
packaging
FDA root cause
Packaging
Action
Consignees notified by first class mail with formal recall letter dated January 07, 2020. The letter identified affected product, stated reason for recall, and requested product to be quarantined. A response form was requested to be returned via email.
Quantity in commerce
500
Distribution
Kentucky. No governmental, Canadian, or Mexican consignees.
Product codes and lots
Code DYNDA1928 Lot Number: 2019032190 Expiration Date 07/31/2020

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0865-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0865-2020Class IITerminatedVoluntary: Firm initiated