Z-0988-2020Class II
Jamshidi Bone Marrow Aspiration Needle Item Number: 00-1103-007-00
Zimmer GmbH / initiated 2019-12-23 / Terminated / root cause: packaging
Zimmer Biomet, Inc. began a recall of Jamshidi Bone Marrow Aspiration Needle Item Number: 00-1103-007-00 on . FDA classified it as Class II and gives the reason as "Potential for the expiration date on the outer label that was applied by Zimmer Biomet to exceed the expiration date on the inner label". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0988-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Jamshidi Bone Marrow Aspiration Needle Item Number: 00-1103-007-00
- Product code
- KNW Instrument, Biopsy
- Reason for recall
- Potential for the expiration date on the outer label that was applied by Zimmer Biomet to exceed the expiration date on the inner label
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Zimmer Biomet issued Urgent Device Recall dated 12/23/19, to Distributors/Hospitals via email/courier stating reason for recall, health risk and action to take: Distributors responsibilities- locating and removing the product in their territory, as well as identifying hospitals and surgeons who have previously used the product. Distributors will return on-hand product to Zimmer Biomet and ensure all of their products are accounted for using the form provided in the letter. Hopsitals: Review this notification and ensure that affected personnel are aware of the contents, including surgeons that utilized this product. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. Questions or concerns call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday.. Alternatively, your questions may be emailed to [email protected]
- Quantity in commerce
- 133 units
- Distribution
- US Nationwide
- Product codes and lots
- Lot Numbers : 62754464, 62904231, 63060634
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0988-2020 | Class II | Terminated | Voluntary: Firm initiated |