Z-1715-2020Class II

Portex LineDraw Arterial Blood Sampling Syringe with Dry Lithium Heparin for Gases and Electrolytes

Abbott Medical / initiated 2020-01-07 / Terminated / root cause: manufacturing process

Smiths Medical ASD, Inc. began a recall of Portex LineDraw Arterial Blood Sampling Syringe with Dry Lithium Heparin for Gases and Electrolytes on . FDA classified it as Class II and gives the reason as "Portex Pro-Vent and LineDraw Arterial Blood Sampling syringes may have broken luers". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1715-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Portex LineDraw Arterial Blood Sampling Syringe with Dry Lithium Heparin for Gases and Electrolytes
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
Portex Pro-Vent and LineDraw Arterial Blood Sampling syringes may have broken luers.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm initiated the recall by letter on January 07, 2020. The "Urgent Medical Device Recall Notice", identified the removal of devices and instructed the customer to quarantine the affected products. The Notice provided the accompanying Recall Response Form to be completed and returned within 10 days of receipt to [email protected]. The Notice details that the affected product must be returned to Stericycle for processing. The Notice instructed the distributor to notify their customers and to provide the customer a copy of the Recall Notice and Response Form. The Notice provided the contact email [email protected] and phone number of 8669188738 for assistance.
Quantity in commerce
720,100 total devices
Distribution
USA: AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI,MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA. Int'l: CA, DE, AU.
Product codes and lots
Portex LineDraw Arterial Blood Sampling Syringe: Item Number Lot Number 4042-2 3848838 4042-2 3848878 4042-2 3856950 4042-2 3859774 4042LH 3856958 4042LH 3859747

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1715-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1715-2020Class IITerminatedVoluntary: Firm initiated
Z-1716-2020Class IITerminatedVoluntary: Firm initiated