Z-1158-2020Class II

cobas infinity core, Material No. 7154003001. The device is a calculator/data processing module, for clinical...

Roche Diagnostics GmbH / initiated 2020-01-07 / Terminated

Roche Diagnostics Operations, Inc. began a recall of cobas infinity core, Material No. 7154003001. The device is a calculator/data processing module, for clinical use on . FDA classified it as Class II and gives the reason as "Alarm flags that have a "<" symbol in front of them are not displayed on the cobas infinity Validation screen. This issue affects only customers that are doing manual validation on the cobas infinity Validation screen". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1158-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
cobas infinity core, Material No. 7154003001. The device is a calculator/data processing module, for clinical use.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Alarm flags that have a "<" symbol in front of them are not displayed on the cobas infinity Validation screen. This issue affects only customers that are doing manual validation on the cobas infinity Validation screen.
FDA root cause
Under Investigation by firm
Action
On 1/7/2020, the firm distributed Urgent Medical Device Correction letters to affected consignees. Customers were asked to do the following: " Be aware of the incorrect display of the alarm flags on the cobas infinity Validation screen if an alarm has a < symbol in front of it and if doing manual validations. " It is not recommended to use the alarm mapping function on the cobas infinity until your system is patched. If your laboratory does use it, please call the Roche Support Network Customer Support Center at 1-800-526-2272 for further information. " Complete the customer response form and return it to the recalling firm. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-2272 if you have questions about the information contained in this UMDC.
Quantity in commerce
26 accounts
Distribution
Distributed to customers in AK, AL, CA, GA, HI, IA, IL, KY, MA, MI, MN, MO, MS, MT, NJ, NY, OR, SC, and TN.
Product codes and lots
All cobas infinity laboratory solution software versions are affected.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1158-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1158-2020Class IITerminatedVoluntary: Firm initiated