Z-1111-2020Class II

Terumo Advanced Perfusion Assembly Pump Guts, 4 Inch, Catalog Number 801805 -

Terumo Cardiovascular Systems Corporation / initiated 2020-01-10 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of Terumo Advanced Perfusion Assembly Pump Guts, 4 Inch, Catalog Number 801805 - Product Usage: is indicated for use for up to 6 hours in the extracorporeal... on . FDA classified it as Class II and gives the reason as "These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1111-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Advanced Perfusion Assembly Pump Guts, 4 Inch, Catalog Number 801805 - Product Usage: is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical proffessional who is experienced in the operation of this or similar equipment. The roller pump is used to move fluids through the cardiopulmonary bypass circuit and can be used for a number of applicatinos including arterial blood pumping, cardioplegia delivery, suction, and venting.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Terumo began notifying customers via telephone to schedule the repairs on 1/2/20. Urgent Medical Device Correction notification letters dated 1/6/20 were sent to customers.
Quantity in commerce
11
Distribution
Worldwide distribution - US Nationwide distribution in the states of CT, FL, LA, MA, MI, NY, and TX. Countries of Hong Kong, India, South Korea, and Taiwan.
Product codes and lots
Serial Numbers: 880559 880560 882808 882809 882810 882811 882812 882813 885541 885542 885543

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1111-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1108-2020Class IITerminatedVoluntary: Firm initiated
Z-1109-2020Class IITerminatedVoluntary: Firm initiated
Z-1110-2020Class IITerminatedVoluntary: Firm initiated
Z-1111-2020Class IITerminatedVoluntary: Firm initiated