Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1098-2020Class II

Telescopic Smoke Evac Pencil, PB, Coated, Product No. SHKTSPCL. The product is a Bovie pencil that is 75...

Stryker Corp. / initiated 2020-01-13 / Terminated

Stryker Instruments Div. of Stryker Corporation began a recall of Telescopic Smoke Evac Pencil, PB, Coated, Product No. SHKTSPCL. The product is a Bovie pencil that is 75 inches long compressed, with a telescoping shaft. UDI... on . FDA classified it as Class II and gives the reason as "Hairline fractures in a component could allow electrical current to arc out of the device, resulting in a burn to the patient and/or user". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1098-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Telescopic Smoke Evac Pencil, PB, Coated, Product No. SHKTSPCL. The product is a Bovie pencil that is 75 inches long compressed, with a telescoping shaft. UDI GTIN 37613327466332
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Hairline fractures in a component could allow electrical current to arc out of the device, resulting in a burn to the patient and/or user.
FDA root cause
Under Investigation by firm
Action
On January 13, 2020, the firm informed customers of the recall via Urgent Medical Device Recall Notification letters. Customers were advised to take the following actions: 1. Immediately check all stock areas and/or operating room storage for affected products. Quarantine and discontinue use of any recalled Stryker SafeAir Smoke Evacuation Pencils or LINA Telescopic Smoke Pencils. 2. Complete the Business Reply Form (BRF) to confirm receipt of Notification and identify how many affected items are currently in your inventory. Please complete and return the BRF even if you dont have any affected product on hand. Return the completed BRF via fax to 8665212762 or email to [email protected]. 3. If you have further distributed this product, please forward the notification to all affected locations. 4. If the BRF for your facility indicates that recalled product is currently on hand, a shipping label will be provided which should be used to return recalled product. Upon receipt of the recalled product, a credit will be applied to your account.
Quantity in commerce
15150
Distribution
Domestic distribution nationwide. Foreign distribution to Netherlands and Australia.
Product codes and lots
1923011, 1923050, 1924004, 1924005, 1924015, 1929038, 1929039, 1929055

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1098-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1097-2020Class IITerminatedVoluntary: Firm initiated
Z-1098-2020Class IITerminatedVoluntary: Firm initiated
Z-1099-2020Class IITerminatedVoluntary: Firm initiated