Teleflex Medical Disposable Aortic Punch, in the following sizes: a) 2.8 mm, REF DP-28K b) 3.6 mm, REF DP-36K...
TELEFLEX MEDICAL INC began a recall of Teleflex Medical Disposable Aortic Punch, in the following sizes: a) 2.8 mm, REF DP-28K b) 3.6 mm, REF DP-36K c) 4.0 mm, REF DP-40K d) 4.4 mm, REF DP-44K e)... on . FDA classified it as Class II and gives the reason as "There is a potential for the device packaging to become compromised; thus, the sterility of the device cannot be guaranteed". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1270-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teleflex Medical
- Product
- Teleflex Medical Disposable Aortic Punch, in the following sizes: a) 2.8 mm, REF DP-28K b) 3.6 mm, REF DP-36K c) 4.0 mm, REF DP-40K d) 4.4 mm, REF DP-44K e) 4.8 mm, REF DP-48K f) 5.2 mm, REF DP-52K g) 5.6 mm, REF DP-56K h) 6.0 mm, REF DP-60K i) 2.8 mm, REF MDP-28K j) 3.6 mm, REF MDP-36K k) 4.0 mm, REF MDP-40K l) 4.4 mm, REF MDP-44K m) 4.8 mm, REF MDP-48K n) 5.2 mm, REF MDP-52K o) 5.6 mm, REF MDP-56K p) 6.0 mm, REF MDP-60K
- Product code
- DWS Instruments, Surgical, Cardiovascular
- Reason for recall
- There is a potential for the device packaging to become compromised; thus, the sterility of the device cannot be guaranteed.
- Root cause tag
- packaging
- FDA root cause
- Under Investigation by firm
- Action
- The firm initiated the recall by letter on 01/13/2020. The actions to be taking be the consignees are as follows: Check inventory. If affected stock is found, discontinue use immediately and quarantine. Complete the acknowledgement form as soon as possible indicating whether you do or do not have stock and fax it to 1-855-419-8507 or email to [email protected]. A customer service representative will contact you with a Return Goods Authorization (RGA) number and provide instructions for the return of products. For additional questions, contact the local sales representative or Customer Service at 1-866-396-2111.
- Quantity in commerce
- 587066 units
- Distribution
- Worldwide
- Product codes and lots
- a) Teleflex Medical Disposable Aortic Punch, 2.8 mm, REF DP-28K Lot Numbers: 74A1501140, 74B1602364, 74E1700597, 74G1602599, 74H1900357, 74L1501549, 74A1601253, 74B1900832, 74E1800886, 74G1700636, 74J1501218, 74L1501977, 74A1601254, 74C1501113, 74E1802643, 74G1700637, 74J1501754, 74L1501978, 74A1602903, 74C1600966, 74E1901667, 74G1700718, 74J1501755, 74L1600387, 74A1602904, 74C1600972, 74F1502019, 74G1900066, 74J1601180, 74L1600389, 74A1603168, 74C1600976, 74F1600350, 74G1900340, 74J1603181, 74L1602761, 74A1603170, 74C1600977, 74F1602476, 74G1902629, 74J1800844, 74L1700357, 74A1701154, 74C1702793, 74F1602477, 74G1903756, 74K1500475, 74L1800444, 74A1802567, 74C1702794, 74F1702731, 74H1502508, 74K1500476, 74L1803409, 74A1903113, 74C1800357, 74F1800411, 74H1502510, 74K1501684, 74L1901787, 74A1903146, 74D1600123, 74F1802291, 74H1601384, 74K1501685, 74M1601318, 74B1501366, 74D1601266, 74F1802292, 74H1602748, 74K1900124, 74M1602080, 74B1501710, 74D1601524, 74G1500954, 74H1700589, 74K1900125, 74M1800915, 74B1600576, 74D1602673, 74G1500955, 74H1700590, 74K1900133, 74M1800916, 74B1600580, 74D1802317, 74G1503008, 74H1701427, 74K1901814, 74M1801620, 74B1601613, 74E1600837, 74G1503009, 74H1701863, 74L1501239, 74M1900014, 74B1601616, 74E1700596, 74G1602598, 74H1800764, 74L1501548 b) Teleflex Medical Disposable Aortic Punch, 3.6 mm, REF DP-36K Lot Numbers: 74A1601251, 74C1501114, 74E1600361, 74F1901768, 74H1602749, 74K1901790, 74A1602049, 74C1501504, 74E1600362, 74G1501254, 74H1802543, 74L1501933, 74A1602205, 74C1600454, 74E1700383, 74G1501584, 74H1802910, 74L1501934, 74A1603169, 74C1600978, 74E1700598, 74G1502240, 74H1900354, 74L1502651, 74A1801612, 74C1702795, 74E1800887, 74G1502247, 74H1901509, 74L1502668, 74A1901239, 74C1702796, 74E1802644, 74G1601150, 74J1501761, 74L1602762, 74A1902997, 74C1800358, 74E1900552, 74G1601151, 74J1501762, 74L1702282, 74A1903148, 74D1503157, 74E1901665, 74G1700719, 74J1702333, 74L1702978, 74B1501361, 74D1600124, 74F1500725, 74G1702124, 74J1702528, 74L1800993, 74B1600581, 74D1600125, 74F1502017, 74G1702125, 74J1703089, 74L1803139, 74B1601617, 74D1802318, 74F1502713, 74G1802091, 74J1900939, 74L1803408, 74B1602365, 74D1900834, 74F1502714, 74G1902121, 74J1901700, 74M1601319, 74B1602372, 74D1901907, 74F1600351, 74H1500734, 74K1500480, 74M1702625, 74B1800721, 74E1500848, 74F1703304, 74H1500735, 74K1501130, 74M1801251, 74B1800722, 74E1501779, 74F1801505, 74H1502517, 74K1501131, 74M1801622, 74B1902990, 74E1502618, 74F1802293, 74H1600757, 74K1502964, 74M1900012, 74B1903028, 74E1503095, 74F1900940, 74H1601385, 74K1600559 c) Teleflex Medical Disposable Aortic Punch, 4.0 mm, REF DP-40K Lot Numbers: 74A1501146, 74A1602181, 74A1702660, 74A1901901, 74B1602046, 74C1503228, 74A1501148, 74A1602360, 74A1800595, 74B1501126, 74B1800506, 74C1503232, 74A1502433, 74A1602361, 74A1800596, 74B1501358, 74B1800723, 74C1600455, 74A1502434, 74A1603161, 74A1801613, 74B1502260, 74B1801756, 74C1600460, 74A1502851, 74A1603162, 74A1802148, 74B1502261, 74B1802580, 74C1600967, 74A1502852, 74A1603163, 74A1802150, 74B1503151, 74C1501122, 74C1600979, 74A1503453, 74A1700378, 74A1802568, 74B1503152, 74C1501123, 74C1600980, 74A1601248, 74A1700766, 74A1900243, 74B1600577, 74C1501505, 74C1702893, 74A1601249, 74A1700767, 74A1900244, 74B1600582, 74C1502259, 74C1702894, 74A1601250, 74A1702659, 74A1901240, 74B1601618, 74C1502260, 74C1800359, 74C1800360, 74E1503093, 74F1601983, 74G1900067, 74H1903474, 74K1602893, 74C1801150, 74E1503099, 74F1602478, 74G1902118, 74J1500711, 74K1602894, 74C1900330, 74E1600838, 74F1602479, 74G1902119, 74J1500712, 74K1700910, 74C1900332, 74E1600839, 74F1603239, 74G1902833, 74J1501214, 74K1701639, 74C1900811, 74E1601257, 74F1603240, 74G1903729, 74J1501215, 74K1702121, 74C1900812, 74E1602244, 74F1700828, 74H1500732, 74J1501766, 74K1702122, 74C1901597, 74E1602245, 74F1701130, 74H1501603, 74J1501767, 74K1702866, 74C1902227, 74E1602810, 74F1701131, 74H1502506, 74J1502404, 74K1802890, 74C1902228, 74E1602811, 74F1702095, 74H1502507, 74J1502405, 74K1900758, 74D1500077, 74E1700599, 74F1702096, 74H1502530, 74J1600424, 74K1900759, 74D1500078, 74E1700600, 74F1702733, 74H1600758, 74J1600425, 74K1901792, 74D1501647, 74E1701203, 74F1702734, 74H1600759, 74J1601181, 74K1902250, 74D1501648, 74E1701234, 74F1703305, 74H1601386, 74J1601737, 74L1500801, 74D1503172, 74E1701236, 74F1800418, 74H1602468, 74J1602579, 74L1500810, 74D1600213, 74E1701755, 74F1801506, 74H1602469, 74J1602677, 74L1501230, 74D1600214, 74E1701758, 74F1802294, 74H1602750, 74J1602678, 74L1501231, 74D1600927, 74E1703220, 74F1802710, 74H1602751, 74J1700737, 74L1501983, 74D1601538, 74E1703236, 74F1802711, 74H1700591, 74J1700738, 74L1502663, 74D1602013, 74E1801602, 74F1900935, 74H1700592, 74J1701829, 74L1600400, 74D1602014, 74E1801603, 74F1900936, 74H1701429, 74J1702529, 74L1600440, 74D1602676, 74E1802105, 74F1900937, 74H1701431, 74J1702530, 74L1601327, 74D1700564, 74E1802106, 74F1901099, 74H1701839, 74J1703090, 74L1601328, 74D1700954, 74E1802645, 74F1901117, 74H1701869, 74J1703091, 74L1602113, 74D1700955, 74E1802984, 74F1901761, 74H1702709, 74J1800845, 74L1602763, 74D1701179, 74E1900554, 74G1500951, 74H1702710, 74J1801321, 74L1602764, 74D1702440, 74E1900593, 74G1501744, 74H1703164, 74J1801322, 74L1700358, 74D1801512, 74E1901447, 74G1502237, 74H1703165, 74J1802104, 74L1700359, 74D1802103, 74E1901448, 74G1503006, 74H1800761, 74J1802105, 74L1701122, 74D1802104, 74E1901663, 74G1600720, 74H1800763, 74J1802736, 74L1701124, 74D1802319, 74E1901664, 74G1601769, 74H1800838, 74J1900237, 74L1701786, 74D1802693, 74E1902715, 74G1602120, 74H1801707, 74J1900956, 74L1702283, 74D1802694, 74E1903772, 74G1602121, 74H1802544, 74J1901697, 74L1702979, 74D1900484, 74F1500753, 74G1602600, 74H1802911, 74J1901698, 74L1800558, 74D1900485, 74F1500754, 74G1700638, 74H1802912, 74J1903043, 74L1800559, 74D1900657, 74F1501151, 74G1700639, 74H1900352, 74K1500471, 74L1800994, 74D1900659, 74F1501152, 74G1701299, 74H1900353, 74K1500472, 74L1800995, 74D1901505, 74F1502021, 74G1701314, 74H1901108, 74K1501121, 74L1801431, 74D1902167, 74F1502022, 74G1702119, 74H1901109, 74K1501122, 74L1803140, 74D1902810, 74F1502217, 74G1702120, 74H1901115, 74K1501680, 74L1900199, 74D1902811, 74F1502710, 74G1800767, 74H1901116, 74K1502261, 74L1900200, 74D1902812, 74F1502711, 74G1800768, 74H1901510, 74K1502262, 74L1900201, 74E1501782, 74F1600352, 74G1801515, 74H1902119, 74K1502270, 74L1900202, 74E1501783, 74F1601101, 74G1802092, 74H1902737, 74K1600560, 74L1900361, 74E1502616, 74F1601102, 74G1802093, 74H1902738, 74K1601326, 74L1900362, 74E1502617, 74F1601325, 74G1802678, 74H1903398, 74K1602034, 74L1900858, 74L1900859, 74M1500805, 74M1601314, 74M1701819, 74M1801624, 74M1900367, 74L1901403, 74M1600393, 74M1602082, 74M1701820, 74M1900010, 74M1901094, 74L1901404, 74M1600394, 74M1701016, 74M1801252, 74M1900017, 74M1901095, 74L1901885, 74M1601313, 74M1701017, 74M1801253, 74M1900366 d) Teleflex Medical Disposable Aortic Punch, 4.4 mm, REF DP-44K Lot Numbers: 74A1501153, 74D1600217, 74E1900553, 74H1502520, 74J1700739, 74L1700360, 74A1502848, 74D1601525, 74E1903148, 74H1502909, 74J1701830, 74L1701125, 74A1602180, 74D1602016, 74E1903763, 74H1600761, 74J1701831, 74L1701787, 74A1800598, 74D1602268, 74F1600354, 74H1600762, 74J1702532, 74L1701788, 74A1801615, 74D1602679, 74F1600355, 74H1601224, 74J1703093, 74L1702285, 74A1801616, 74D1700951, 74F1601104, 74H1601387, 74J1800846, 74L1702981, 74A1802152, 74D1701786, 74F1601140, 74H1602470, 74J1801323, 74L1702982, 74A1802569, 74D1702441, 74F1601984, 74H1602471, 74J1802106, 74L1800560, 74A1900362, 74D1802105, 74F1601985, 74H1700594, 74J1900625, 74L1800561, 74A1901242, 74D1802106, 74F1603241, 74H1700595, 74J1901696, 74L1800997, 74A1901899, 74D1900483, 74F1603242, 74H1701435, 74J1903042, 74L1800998, 74A1902221, 74D1900658, 74F1702097, 74H1701864, 74K1500483, 74L1801771, 74A1902222, 74D1901506, 74F1702735, 74H1701865, 74K1501132, 74L1803136, 74A1902354, 74D1902168, 74F1800868, 74H1702712, 74K1501688, 74L1803261, 74B1501363, 74E1500841, 74F1801508, 74H1702713, 74K1600561, 74L1901191, 74B1502275, 74E1500842, 74F1802712, 74H1703166, 74K1601328, 74L1901200, 74B1601620, 74E1501992, 74F1900938, 74H1703282, 74K1601329, 74L1901201, 74B1602048, 74E1503098, 74G1502238, 74H1800765, 74K1700912, 74L1901407, 74B1602049, 74E1600364, 74G1503014, 74H1800840, 74K1700913, 74M1600396, 74B1800508, 74E1601723, 74G1600917, 74H1800841, 74K1701641, 74M1601316, 74B1800725, 74E1601724, 74G1601771, 74H1801709, 74K1701642, 74M1601317, 74B1800726, 74E1602248, 74G1601772, 74H1801710, 74K1702868, 74M1602089, 74B1801757, 74E1602812, 74G1602122, 74H1802546, 74K1702869, 74M1602090, 74B1802581, 74E1700602, 74G1602123, 74H1802914, 74K1900131, 74M1702628, 74B1802582, 74E1701406, 74G1602673, 74H1900355, 74K1901789, 74M1702629, 74C1501117, 74E1702791, 74G1602718, 74H1901113, 74K1901813, 74M1900011, 74C1501506, 74E1702792, 74G1700640, 74H1901511, 74L1501242, 74M1900372, 74C1600973, 74E1702796, 74G1701317, 74H1902120, 74L1502469, 74M1900689, 74C1600975, 74E1702797, 74G1701318, 74H1903403, 74L1600402, 74M1901097, 74C1702792, 74E1800889, 74G1801516, 74J1600426, 74L1600403, 74C1902226, 74E1801605, 74G1802681, 74J1600427, 74L1601647, 74D1500076, 74E1801606, 74G1900338, 74J1601203, 74L1601648, 74D1501516, 74E1802108, 74G1902123, 74J1602434, 74L1602115, 74D1501645, 74E1802109, 74G1902633, 74J1603182, 74L1602766, 74D1600216, 74E1802987, 74G1903757, 74J1603183, 74L1602767 e) Teleflex Medical Disposable Aortic Punch, 4.8 mm, REF DP-48K Lot Numbers: 74A1502850, 74A1801617, 74B1502262, 74B1802583, 74C1801152, 74D1600931, 74A1602182, 74A1802153, 74B1503158, 74B1901661, 74C1900329, 74D1601537, 74A1602901, 74A1901238, 74B1601609, 74C1501119, 74C1900618, 74D1602017, 74A1603164, 74A1901900, 74B1800505, 74C1600462, 74D1600226, 74D1602018, 74A1700763, 74A1902355, 74B1800719, 74C1600974, 74D1600227, 74D1700939, 74A1800600, 74B1501128, 74B1800727, 74C1600982, 74D1600930, 74D1700943, 74D1702442, 74F1600357, 74G1601774, 74H1701867, 74J1703095, 74K1900760, 74D1702443, 74F1600358, 74G1602124, 74H1701868, 74J1800847, 74K1901336, 74D1802107, 74F1601105, 74G1602719, 74H1702714, 74J1801324, 74L1500803, 74D1802321, 74F1601137, 74G1602720, 74H1800767, 74J1802108, 74L1600443, 74D1901508, 74F1601986, 74G1701319, 74H1800843, 74J1901706, 74L1602116, 74D1902169, 74F1602009, 74G1701320, 74H1800844, 74J1903044, 74L1701123, 74E1600842, 74F1603243, 74G1800770, 74H1802548, 74K1501123, 74L1701789, 74E1602249, 74F1702099, 74G1801517, 74H1802549, 74K1501124, 74L1702286, 74E1602813, 74F1702736, 74G1900337, 74H1802915, 74K1600562, 74L1800443, 74E1701837, 74F1702737, 74G1900339, 74H1900356, 74K1600563, 74L1800999, 74E1701838, 74F1703306, 74G1902120, 74H1901512, 74K1602037, 74L1803265, 74E1801607, 74F1800866, 74G1903728, 74J1501216, 74K1602038, 74M1600397, 74E1801608, 74F1801510, 74G1903758, 74J1501759, 74K1700915, 74M1602092, 74E1802646, 74F1802296, 74H1601389, 74J1600428, 74K1701643, 74M1701020, 74E1802647, 74F1900939, 74H1601390, 74J1601183, 74K1701644, 74M1702630, 74E1902716, 74F1901122, 74H1602472, 74J1601783, 74K1702870, 74M1702631, 74E1902717, 74F1902895, 74H1602752, 74J1602085, 74K1702871, 74M1800913, 74F1502712, 74G1501749, 74H1602753, 74J1700740, 74K1900128, 74M1900368, 74F1503522, 74G1601773, 74H1700596, 74J1701832, 74K1900129, 74M1901096 f) Teleflex Medical Disposable Aortic Punch, 5.2 mm, REF DP-52K Lot Numbers: 74A1601258, 74D1700800, 74F1602482, 74H1600764, 74K1602778, 74M1800914, 74A1700762, 74D1700820, 74F1701133, 74H1701443, 74K1700916, 74A1800602, 74D1701167, 74F1701134, 74H1702715, 74K1702872, 74A1903114, 74D1802322, 74F1800869, 74H1800846, 74K1900132, 74B1501364, 74D1802323, 74F1800870, 74H1801712, 74K1900761, 74B1503155, 74E1600843, 74F1801512, 74H1801713, 74K1901795, 74B1600584, 74E1602250, 74F1901098, 74H1802550, 74K1902045, 74B1602050, 74E1602814, 74G1500952, 74J1601733, 74K1902545, 74B1800510, 74E1700899, 74G1700720, 74J1602581, 74L1501241, 74B1800729, 74E1701840, 74G1900341, 74J1701833, 74L1602118, 74B1802584, 74E1701841, 74G1902122, 74J1802110, 74L1700361, 74B1900830, 74E1800897, 74G1902834, 74J1900239, 74L1800555, 74C1501507, 74E1900596, 74G1903730, 74J1901704, 74L1803267, 74C1902229, 74F1500755, 74G1903755, 74K1501689, 74M1600398, 74D1602269, 74F1601106, 74H1600763, 74K1600564, 74M1601320 g) Teleflex Medical Disposable Aortic Punch, 5.6 mm, REF DP-56K Lot Numbers: 74A1501142, 74B1600585, 74D1501515, 74F1500724, 74H1702716, 74K1902252, 74A1601256, 74B1601614, 74D1502568, 74F1502020, 74J1702533, 74L1501976, 74A1602204, 74B1601622, 74E1500845, 74F1702103, 74K1500481, 74L1502672, 74A1603172, 74B1602367, 74E1502580, 74F1800871, 74K1501683, 74M1900365, 74A1800603, 74B1800720, 74E1602815, 74F1801514, 74K1502970, 74M1900840, 74B1500704, 74C1501508, 74E1701843, 74F1801515, 74K1602779, 74M1901098, 74B1502274, 74C1600968, 74E1701844, 74F1802297, 74K1900126, 74M1901275, 74B1600578, 74D1500075, 74E1800898, 74G1902124, 74K1900127 h) Teleflex Medical Disposable Aortic Punch, 6.0 mm, REF DP-60K Lot Numbers: 74A1500172, 74E1502579, 74M1602094, 74E1800901, 74G1902125, 74D1501639, 74A1602907, 74H1700597, 74E1800902, 74K1901815, 74E1500846, 74E1602816, 74H1702717, 74G1900342 i) Teleflex Medical Disposable Aortic Punch, 2.8 mm, REF MDP-28K Lot Numbers: 74A1502835, 74A1902357, 74E1701845, 74F1902101, 74H1801714, 74A1601252, 74B1501360, 74E1901451, 74F1902891, 74H1802551, 74K1901816, 74A1602906, 74B1601615, 74F1500722, 74F1902892, 74J1702534, 74L1702287, 74A1800642, 74B1602374, 74F1501155, 74H1500733, 74J1901705, 74A1802572, 74B1801755, 74F1800872, 74H1702718, 74K1800082, 74A1901243, 74D1700801, 74F1802298, 74H1702719, 74K1900135 j) Teleflex Medical Disposable Aortic Punch, 3.6 mm, REF MDP-36K Lot Numbers: 74A1602184, 74B1602375, 74F1500984, 74H1601392, 74J1802111, 74L1702288, 74A1602185, 74C1702803, 74F1703309, 74H1701444, 74K1500473, 74L1702289, 74A1802573, 74C1702804, 74F1800873, 74H1702720, 74K1602780, 74L1901405, 74A1901218, 74C1802507, 74F1901968, 74H1801715, 74K1701646, 74L1901411, 74A1902223, 74C1802508, 74F1902108, 74H1802552, 74K1900134, 74M1901099, 74B1501261, 74E1600363, 74G1500956, 74H1802553, 74L1500804, 74B1600586, 74E1701846, 74G1502248, 74H1902743, 74L1500805, 74B1601623, 74E1701847, 74H1502513, 74J1500701, 74L1501981, 74B1602368, 74E1902721, 74H1502514, 74J1702535, 74L1501982 k) Teleflex Medical Disposable Aortic Punch, 4.0 mm, REF MDP-40K Lot Numbers: 74A1503439, 74C1802506, 74F1502715, 74G1602722, 74J1502407, 74K1801695, 74A1601257, 74C1900617, 74F1503526, 74G1800771, 74J1601185, 74L1500807, 74A1602908, 74D1600957, 74F1503527, 74G1801518, 74J1601186, 74L1500808, 74A1603166, 74D1600958, 74F1600359, 74G1802094, 74J1700494, 74L1501935, 74A1700734, 74D1601536, 74F1601332, 74G1902832, 74J1700742, 74L1501936, 74A1701848, 74D1602680, 74F1601333, 74H1500736, 74J1802112, 74L1602768, 74A1802574, 74D1700567, 74F1603244, 74H1502518, 74J1901699, 74L1602769, 74B1501131, 74D1700568, 74F1702104, 74H1502519, 74K1500478, 74L1701126, 74B1501365, 74D1800226, 74F1702105, 74H1602475, 74K1500479, 74L1702983, 74B1503154, 74D1800227, 74F1702738, 74H1602476, 74K1502266, 74L1900759, 74B1600587, 74E1502619, 74F1702739, 74H1701446, 74K1502268, 74L1900860, 74B1601624, 74E1503096, 74F1703310, 74H1702721, 74K1600565, 74M1602095, 74B1602369, 74E1601726, 74F1902081, 74H1902739, 74K1601332, 74M1701824, 74B1802585, 74E1602817, 74F1902177, 74H1902740, 74K1602781, 74M1701825, 74C1600456, 74E1602818, 74F1902178, 74H1902741, 74K1602782, 74M1900369, 74C1600984, 74F1500759, 74G1501993, 74H1903402, 74K1701647, 74M1901093, 74C1600985, 74F1501153, 74G1501994, 74J1501756, 74K1702867, 74C1802505, 74F1502018, 74G1602721, 74J1502406, 74K1800961 l) Teleflex Medical Disposable Aortic Punch, 4.4 mm, REF MDP-44K Lot Numbers: 74A1502836, 74B1902383, 74D1801513, 74F1802299, 74H1702722, 74K1702124, 74A1601255, 74C1503233, 74D1801514, 74F1901118, 74H1801716, 74K1800962, 74A1602179, 74C1702806, 74D1900660, 74G1500953, 74H1900358, 74K1801699, 74A1700760, 74C1801154, 74E1500844, 74G1601152, 74H1901513, 74L1501916, 74A1700761, 74C1900328, 74E1601486, 74G1602723, 74H1902742, 74L1502660, 74A1800643, 74C1901596, 74E1602252, 74G1700721, 74J1603175, 74L1601649, 74A1800644, 74D1601526, 74E1602819, 74G1802095, 74J1603176, 74L1700363, 74A1801620, 74D1601535, 74E1701849, 74H1502515, 74J1700743, 74L1700364, 74A1802575, 74D1602681, 74E1900594, 74H1601394, 74J1703098, 74L1702290, 74B1501359, 74D1700916, 74F1600360, 74H1601395, 74J1800849, 74L1702984, 74B1600575, 74D1800228, 74F1601138, 74H1701448, 74J1901702, 74L1801776, 74B1600588, 74D1800229, 74F1703312, 74H1701449, 74K1602783, 74L1900363, 74L1900760, 74M1602086, 74M1701826, 74L1901239, 74M1701022, 74M1900013 m) Teleflex Medical Disposable Aortic Punch, 4.8 mm, REF MDP-48K Lot Numbers: 74A1602178, 74C1902230, 74F1600361, 74H1502516, 74J1800850, 74K1801700, 74A1801621, 74D1503161, 74F1600362, 74H1601396, 74J1800851, 74K1801701, 74A1802576, 74D1602682, 74F1601982, 74H1602477, 74J1802113, 74L1501243, 74A1802577, 74D1700565, 74F1602484, 74H1603433, 74J1802740, 74L1602770, 74A1900255, 74D1700566, 74F1700830, 74H1701450, 74J1900241, 74L1700365, 74A1902358, 74D1800230, 74F1800874, 74H1701840, 74J1901703, 74L1701127, 74B1600589, 74D1800231, 74F1802300, 74H1802554, 74K1600566, 74L1803269, 74B1800512, 74E1501776, 74G1600722, 74H1901514, 74K1600567, 74L1901406, 74B1800730, 74E1601487, 74G1602125, 74H1903400, 74K1601333, 74M1601321, 74B1802586, 74E1601494, 74G1802096, 74J1501758, 74K1601334, 74M1602097, 74C1501509, 74E1703226, 74G1802684, 74J1600431, 74K1602784, 74M1701023, 74C1801155, 74E1801610, 74G1902626, 74J1601187, 74K1700917, 74M1701827, 74C1802503, 74E1802702, 74G1902835, 74J1601188, 74K1702873, 74M1701828, 74C1802504, 74E1901452, 74G1903731, 74J1702563, 74K1800963 n) Teleflex Medical Disposable Aortic Punch, 5.2 mm, REF MDP-52K Lot Numbers: 74A1603171, 74C1502258, 74D1902815, 74F1902890, 74H1701451, 74L1702291, 74A1603177, 74C1600457, 74F1500987, 74F1902893, 74H1701452, 74M1600399, 74A1801622, 74C1600464, 74F1600363, 74G1501260, 74H1801717, 74M1900015, 74A1900570, 74D1602270, 74F1702740, 74G1503013, 74J1502610, 74M1900690, 74A1903115, 74D1700574, 74F1702741, 74G1801519, 74K1502265, 74M1901101, 74B1501133, 74D1700575, 74F1800875, 74H1502512, 74K1602785, 74C1501510, 74D1800232, 74F1902113, 74H1601397, 74L1500806 o) Teleflex Medical Disposable Aortic Punch, 5.6 mm, REF MDP-56K Lot Numbers: 74A1502837, 74A1601260, 74A1602183, 74B1600579, 74C1600458, 74D1901016 p) Teleflex Medical Disposable Aortic Punch, 6.0 mm, REF MDP-60K Lot Numbers: 74A1500226, 74B1503153, 74C1900613, 74F1602485, 74H1500728, 74J1602582, 74A1601259, 74B1602370, 74D1500074, 74F1702742, 74H1502910, 74J1901701, 74A1700732, 74C1502256, 74D1700569, 74F1800876, 74H1601398, 74K1800083, 74A1801623, 74C1600459, 74D1801515, 74G1501745, 74H1702723, 74K1901793, 74A1802579, 74C1600970, 74E1700603, 74G1801520, 74H1702724, 74L1601650, 74A1901241, 74C1800374, 74F1500723, 74G1902631, 74H1801718, 74L1700366, 74B1501262, 74C1802502, 74F1502701, 74G1902818, 74H1802917
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1269-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1270-2020 | Class II | Terminated | Voluntary: Firm initiated |